Close Btn

Select Your Regional site

Close

ESR measurement condition ** Magnetic field sweep width **

ER230008E

Optimization of ESR measurement condition

In order to measure ESR under optimal conditions, it is important to consider the following items. For more detailed information about  (1) - (3)  , please refer to the application notes ((1) : ER220001、ER220005、(2) : ER220003、(3) : ER220004).

(1) Microwave power
(2) Magnetic modulation width
(3) Sweep time and Time constant
(4) Magnetic field sweep width

Line width and range of ESR signal in various samples

The sample condition can affect the line shape of ESR signal. Generally, the line width of the ESR signal for solid samples is wider than that for solution samples. Table 1 shows various samples and their respective ranges of ESR signals. Radicals in solution exhibit fast molecular motion, causing their anisotropy to average out to produce a symmetrical and sharp ESR signal. Organic free radicals in solution exhibit a narrow line width ranging from 5 μT to 50 μT. Typically, organic radicals in solid have a line width of 0.1 mT or more. In solid samples of polycrystalline, asymmetric signals due to anisotropy appear because ESR signals of radicals with various orientations overlap with respect to the magnetic field. The molecular motion of radicals in highly viscous solutions is suppressed, leading to an ESR signal line shape that falls intermediate between that of a solution and a solid. 

Table 1. Comparison of ESR signals ranges in various samples.

(①、②:Solution、③〜⑥:Solid(powder)、⑦:Gas)

Number Sample ESR signal area (mT) Application note
① Perylene 4 ER220004E
② TEMPOL 5 ER220001E
③ Carbon 3 ER220005E
④ Ultramarine 25 ER200002E
⑤ Copper sulfate 100 ER220001E
⑥ Chromium oxide (Ⅲ) 600 ER200003E
⑦ Oxygen > 1000 ER230006E

Setting magnetic field sweep width

The  value of the magnetic field at which the ESR signal is observed is determined by the resonance condition expression: hν = gμBH. The magnetic field sweep widths for various samples are provided below. If the magnetic field sweep width becomes too wide for the observed signal, it can lead to a deterioration in the signal response.

Sample of organic radicals (N、O、Si etc.)

335 / 325 ± 10 mT (Resonance frequency:around 9.4 / 9.1 GHz )

Sample of metal complex

300 ± 250 mT

 
ν : Resonance frequency、H : Resonance  magnetic field、h : Plank constant、 μB : Bohr magneton、g : g-factor

* Please also note that when using attachments such as a variable temperature controller,  the resonance frequency and the resonance magnetic field will change.

 

Solutions by field

Related products

Close
Notice

Are you a medical professional or personnel engaged in medical care?

Yes

No

Please be reminded that these pages are not intended to provide the general public with information about the products.

JEOL Instrument Basics

Easy explanation about mechanisms and
applications of JEOL products

Contacts

JEOL provides a variety of support services to ensure that our customers can use our products with peace of mind.
Please feel free to contact us.