【DISCONTINUED】JEM-3200FS Field Emission Energy Filter Electron Microscope

DISCONTINUED

JEM-3200FS Field Emission Energy Filter Electron Microscope

This product is no longer available.

If you would like to know the latest information about your preferred product or to find out more about alternatives, please click on the link below. We hope you will continue to use our products.

The JEM-3200FS Field Emission Electron Microscope is equipped with a field emission electron gun of 300 kV accelerating voltage and an in-column energy filter which shows high performance. This electron microscope offers solutions for various biological and materials researches.

Features

In-column energy filter (Omega filter)

The in-column energy filter, built into the imaging lens system, makes possible a magnification range comparable to those obtained with conventional electron microscopes, and makes energy filtered images and energy loss spectra easy to acquire. The electron optical system of the energy filter (omega filter) is designed to minimize the distortion of images, and furthermore, residual distortion is finally eliminated in the factory before shipping. 
The optional high sensitivity camera system and optional image processing system make possible the construction of an element mapping system and energy loss spectroscopy system.

New control systems

Systemizing basic functions of electron microscopy, such as the electron gun, electron optical system, goniometer, and evacuation system realizes a highly stabilized, high performance control system. 

The use of Windows®*1 for the image screens and user interface makes possible the programming of  operations and the central control of attachments.

New goniometer

The new systemized goniometer with piezoelectric driving elements shows much improvement in specimen shift at high magnifications*2.

  • Notices:Windows is a registered trademark of Microsoft Corporation in the United States and other countries.

  • Only the piezoelectric elements are included in the basic unit. The power supply for the piezoelectric elements is optional.

Specifications

Configuration※1 Ultrahigh resolution
(UHR)
High resolution
(HR)
High specimen tilt
(HT)
High contrast
(HC)
Resolution(nm)
 Point
 Lattice
0.17
0.1
0.19
0.1
0.21
0.1
0.26
0.14
Energy Resolution 0.9 eV(zero loss FWHM)
Accelerating Voltage 300kV,200kV,100kV※2
Step size
 Energy shift
100V minimum
3,000 V maximum (0.2 V steps)
Electron source
Emitter ZrO/W(100) Schottky emitter
Brightness ≧7×108 A/cm2・sr
Vacuum About 3×10-8Pa
Probe current 0.5nA for 1nm probe
Power Stability
Acc. voltage 2×10-6/min
OL current 1×10-6/min
Filter lens current 1×10-6/min
Objective Lens
Focal length 2.7 mm 3.0 mm 3.5 mm 3.9 mm
Spherical aberration 0.6 mm 1.1 mm 1.4 mm 3.2 mm
Chromatic aberration 1.5 mm 1.8 mm 2.2 mm 3.0 mm
Minimum step 1.0 nm 1.4 nm 1.5 nm 4.1 nm
Spot Size (nm φ in diameter)
TEM mode 2 to 5 2 to 5 7 to 30
EDS mode 0.4 to 1.6 0.5 to 2.4 4 to 20
NBD mode 0.4 to 1.6 0.5 to 2.4 -
CBD mode 0.4 to 1.6 0.5 to 2.4 -
Parameters for Convergent beam Diffraction
Convergent angle(2α) 1.5 to 20 mrad -
Acceptance angle ±10 ° -
Magnification
MAG mode ×2,500 to 1,500,000 ×2,000 to 1,200,000 ×1,500 to 1,200,000
LOW MAG mode ×100 to 3,000
SA MAG mode ×8,000 to 600,000 ×6,000 to 500,000 ×5,000 to 500,000
Field of View Slit width at 10 eV 60 mm (on film surface)
Slit width at 2 eV 25 mm (on film surface)
Camera Length
SA diffraction 200 to 2,000 mm 250 to 2,500 mm 400 to 3,000 mm
EELS Dispersion
On energy selection slit 0.85 μm/eV at 300 kV
On film surface 100 to 220 μm/eV at 300kV
Specimen Chamber
Specimen travel range (X,Y;Z)(mm) X,Y : 2
Z : 0.2
X,Y : 2
Z : 0.4
Specimen tilting (X/Y) ±25°/±25°※3 ±35°/±30°※3 ±42°/±30°※3 ±38°/±30°※3
EDS※4
Solid angle 0.13 sr 0.09 sr
Take-off angle 25° 20°
  • Specify either configuration (UHR, HR, HT or HC) when ordering the JEM-3200FS.

  • 100 kV and 200 kV become possible when using optional short circuit switches for the accelerating tube.

  • When the Specimen Tilting Holder (EM-31630) is used.

  • An optional EDS is needed. Specifications apply to a 30 mm2 detector.

Application

Application JEM-3200FS

Gallery

More Info

JEOL Instrument Basics

Easy explanation about mechanisms and
applications of JEOL products

Close
Notice

Are you a medical professional or personnel engaged in medical care?

Yes

No

Please be reminded that these pages are not intended to provide the general public with information about the products.

Contacts

JEOL provides a variety of support services to ensure that our customers can use our products with peace of mind.
Please feel free to contact us.