Close Btn

Select Your Regional site

Close

Residual solvent analysis in cosmetic ingredient using single QMS mode and alternative carrier gases in GC/MS/MS

MSTips No. 466

Introduction

Fig. 1 JMS-TQ4000GC UltraQuad™ TQ

The gas chromatograph triple quadrupole mass spectrometer JMS-TQ4000GC UltraQuad™ TQ (Fig. 1) can be used as QMS by single QMS mode in addition to qualitative and quantitative analysis of trace components by MS/MS mode. We have already reported quantitative analysis in water samples and qualitative analysis in material samples in MSTips No. 443 and 450. In this time, we report the analysis of residual solvents in cosmetic ingredient using the single QMS mode. Residual solvents in cosmetic ingredient, like residual solvents in pharmaceuticals, are classified as Class 1-3 solvents by the International Council for Harmonization of Technical Requirements for Pharmaceuticals for Human Use (ICH). Class 1 solvents: Solvents to be avoided in the manufacture of pharmaceuticals. Class 2 solvents: Solvents to be limited in pharmaceuticals. Class 3 solvents: Solvents with low toxic potential[1]. In this application note, we report the analysis of residual solvents in ethanol solution using nitrogen gas as an alternative carrier gas for seven solvent components classified as Class 1 and 2, as well as the quantitative analysis of 1,4-Dioxane in cosmetic ingredient.

Method

The measurement was performed using a MS-62071STRAP trap-type HS and GC triple quadrupole mass spectrometer JMS-TQ4000GC UltraQuad™ TQ. The measurement conditions are shown in Table 1, and the target components and their SIM monitoring ion are shown in Table 2.

Table 1 Measurement condition

Sample
Sample 1 Standard 7 mix in EtOH
(Mixture of Methyl alcohol, Methylene chloride, Trichloromethane, Benzene, 1,4-Dioxane, Toluene and Xylene including Ethylbenzene. Concentration levels for each compound range from 0.2-100 ppm.)
Sample 2 Cosmetic ingredient including 1,4-Dioxane at about 1 ppm
HS condition
Sample Volume Sample 1: 2 μL, Sample 2: about 1g with 1,4-Dioxane 4 μL and d8-Dioxane 4 μL
Sample Temp. Sample 1: 100°C, Sample 2: 80°C
Sampling mode Loop mode
Heating time Sample 1: 60 min, Sample 2: 30 min
GC condition
Column Rxi-624 (30 m length, 0.25 mm i.d., 1.4 μm film thickness)
Inlet Split/Splitless
Inlet Temp. 200°C
Flow He carrier gas 1 mL/min, Constant flow
N2 carrier gas 1 mL/min, Constant flow
Injection Mode Split (50 :1)
Oven Program 35°C (5 min) → 10°C/min → 250°C (5 min)
MS condition
Ion Source Temp. 200°C
Interface Temp. 200°C
Ionization Mode EI+, 70 eV (He carrier gas), 20 eV (N2 carrier gas)
Measurement Mode SIM

Table 2 Monitoring ion of SIM

Component Quantitative ion Reference ion 1 Reference ion 2
Methyl Alcohol 31 29 32
Methylene chloride 84 49 86
Trichloromethane 83 85 118
Benzene 78 77 52
d8-Dioxane 96 64 46
1,4-Dioxane 88 58 43
Toluene 91 92 65
Ethylbenzene 91 106 65
m,p-Xylene 91 106 65
o-Xylene 91 106 65

Result

⚫︎Measurement result of sample 1

The observed EIC peaks of each component by He carrier gas and N2 carrier gas are shown in Fig. 2. All target components were observed by both carrier gases. On the other hand, when focusing on methyl alcohol, the EIC peak of m/z 29 is buried in the baseline for the N2 carrier gas. It is supposed that baseline was higher than that of the He carrier gas by stable isotope(15N) of N2. The comparison result of S/N value is shown in Table 3. Generally, it is said that the sensitivity of N2 carrier gas is about 1/10 lower than that of He carrier gas. However, for this measurement, the S/N values of 8 out of 9 components were within 1/5. It is assumed that the ionization energy of 20 eV suppressed the generation of nitrogen ion, which is a factor in sensitivity reduction, and as a result, good sensitivity was obtained.

 

Fig. 2

Fig. 2 The observed EIC peaks of each component by He carrier gas and N2 carrier gas

Table 3 Comparison S/N value by He carrier gas with S/N value by N2 carrier gas

Component S/N value by He carrier gas S/N value by N2 carrier gas
Methyl Alcohol 105022.7 24079.0
Methylene chloride 14312.2 10044.2
Trichloromethane 7131.1 5385.4
Benzene 2288.3 1065.5
1,4-Dioxane 3098.7 913.9
Toluene 1365053.9 256462.1
Ethylbenzene 33704.9 4821.1
m,p-Xylene 99506.4 33379.6
o-Xylene 20814.7 9726.1

⚫︎Measurement result of sample 2

The quantitative value of 1,4-Dioxane was calculated by the standard addition method. The ratio of peak area value of 1,4-Dioxane and d8-Dioxane by both carrier gases are shown in Table 4. The EIC peaks and calibration curves of 1,4-Dioxane obtained from cosmetic ingredient by both carrier gases are shown in Fig. 3. The EIC peaks of m/z 88 of 1,4-Dioxane was observed with a good peak shape from the cosmetic ingredient, and the calibration curve showed a linearity of more than 0.99. The quantitative value of 1,4-Dioxane was 1.13 ppm by He carrier gas and 1.11 ppm by N2 carrier gas. These values were close to the indicated concentration.

Table 4 The ratio of peak area value of 1,4-Dioxane to peak area value of d8-Dioxane

Added concentration of
1,4-Dioxane
Ratio of peak area value
by He
Ratio of peak area value
by N2
0 ppm 0.645 0.620
250 ppm 1.268 1.259
500 ppm 1.822 1.786
Fig. 3

Fig. 3 Calibration curve by standard addition method and the peak of 1,4-Dioxane in cosmetic ingredient

Conclusion

It was shown that residual solvent analysis is possible using the single QMS mode of the JMS-TQ4000GC. Regarding the SIM measurement by N2 carrier gas, it was confirmed that the peak intensity was sufficient for residual solvent analysis. Furthermore, the JMS-TQ4000GC can perform single QMS mode as well as MS/MS mode with a single instrument. Thus, it is possible to select the measurement mode according to the amount of matrix in the sample, making it applicable to a wider range of application fields.

Acknowledgments

We would like to thank Ms. Yasuda, Ms. Kobayashi and Ms. Yamaguchi of KOSÉ Corporation for providing samples for the preparation of this application note.

References

[1] Japanese Pharmacopoeia 18th Edition (2.46 Residual Solvents)

 

Solutions by field

Related products

JMS-TQ4000GC Triple Quadrupole Mass Spectrometer

Close
Notice

Are you a medical professional or personnel engaged in medical care?

Yes

No

Please be reminded that these pages are not intended to provide the general public with information about the products.

JEOL Instrument Basics

Easy explanation about mechanisms and
applications of JEOL products

Contacts

JEOL provides a variety of support services to ensure that our customers can use our products with peace of mind.
Please feel free to contact us.