【DISCONTINUED】IB-09060CIS Cryo Ion Slicer™

DISCONTINUED

IB-09060CIS Cryo Ion Slicer™

This product is no longer available.

If you would like to know the latest information about your preferred product or to find out more about alternatives, please click on the link below. We hope you will continue to use our products.

Features

For easy preparation of TEM thin film samples of heat sensitive materials

  • Minimizes thermal damage during ion irradiation using liquid nitrogen

  • Long cryo retention time to fully eliminate damage

  • Specimen exchange with liquid nitrogen injected

  • Maximum sample size: 2.8 mm (L) x 1.0 mm (W) x 0.1 mm (T)

Specifications

IB-09060CIS
Ion accelerating voltage 1 to 8kV
Tilt angle Up to 6°(0.1°step)
Ion-beam diameter 500 µm or more (full width at half maximum)
Milling speed 5 µm/min(accelerating voltage: 8 kV, Si equivalent value)
Gas Argon
Maximum specimen size 2.8mm(L)×1.0 mm(W)×0.1mm(T)
試料冷却 Specimen-holdercooling temperature -120℃
Time for cooling attainment 1h
Cooling-retention time 8h
Specimen take-out time 30min
Pressure measurement Penning gauge
Main vacuum pump Turbo-molecular pump
CCD Camera Built-in
Dimensions and masses Basic unit 500mm(W)×600mm(D)×542mm(H)、70kg
Rotary pump 150mm(W)×427mm(D)×230.5mm(H)、16kg
Display 326mm(W)×173mm(D)×380mm(H)、3.7kg

Installation Requirements

IB-09060CIS
Power Single-phase AC 100 to 120 V±10 %, 50/60 Hz,0.5 to 0.6 kVA
Grounding 100Ω or less
Argon gas Pressure: 0.15±0.05 MPa (1.0 to 2.0 kg/cm2)Purity: 99.9999% or more (argon gas, gas cylinder and regulator should beprovided by the customer)
Hose joint: JIS B0203, Rc 1/4
Room temperature 20 to 25℃
Humidity 60% or less
  • Specifications subject to change without prior notice.

Application

Application IB-09060CIS

More Info

JEOL Instrument Basics

Easy explanation about mechanisms and
applications of JEOL products

Close
Notice

Are you a medical professional or personnel engaged in medical care?

Yes

No

Please be reminded that these pages are not intended to provide the general public with information about the products.

Contacts

JEOL provides a variety of support services to ensure that our customers can use our products with peace of mind.
Please feel free to contact us.