【DISCONTINUED】EM-09100IS Ion Slicer™

DISCONTINUED

EM-09100IS Ion Slicer™

This product is no longer available.

If you would like to know the latest information about your preferred product or to find out more about alternatives, please click on the link below. We hope you will continue to use our products.

Features

Innovative Specimen Preparation Method for TEM / STEM / SEM / EPMA / AUGER

 

The Ion Slicer can prepare thin-film specimens without solvents or chemicals and requires no prior treatment of the specimen other than rectangular slicing (no disc grinding or dimple grinding).

The Ion Slicer prepares thin-film specimens faster and easier than conventional preparation tools. A low-energy, low-angle Ar ion beam irradiates the specimen while a thin shield belt allows low-angle irradiation of the Ar ion beam (from 0° to 6°), drastically reducing ion-beam irradiation damage to the specimen. The result is a high-quality thin film with few sputtering artifacts--even in soft materials. The Ion Slicer can efficiently prepare thin films from specimens having different compositions, even those having porous composites.

Highlights include:

  • High quality TEM pre-treatment

  • Fast preparation

  • No complicated pre-treatments

  • Minimal surface damage

Specifications

EM-09100IS
Ion accelerating voltage 1 to 8 kV
Tilt angle Up to 6° (0.1°step)
Beam diameter 500 μm (FWHM)
Milling rate 5m/min (8 kV, silicon)
Gas for beam irradiation Argon
Recommended specimen size 2.8 mm (in length) 0.5 mm (in width) 0.1 mm (in thickness)
Pressure measurement Penning gauge
Main evacuation system Turbo-molecular pump
CCD camera Built-in
Dimension and weight Main console 500 mm (W) ×600 mm (D) ×542 mm (H)、63 kg
Rotary pump 150 mm (W) ×427 mm (D) ×230.5 mm (H)、16 kg
LCD monitor 326 mm (W) ×173 mm (D) ×380 mm (H)、3.7 kg

Installation Requirements

EM-09100IS
Power Single phase, AC100 to 120V, 50/60 Hz, 500 to 600VA
Grounding terminal 100 ohms or less
Argon gas Pressure: 0.15±0.05MPa(1.0 ∼ 2.0kg/cm2)
Argon gas flow rate: About 0.2 ccm
Purity: 99.9999% or more
Hose joint: JIS B0203 Rc 1/8
*Argon gas (including piping and gas container) and a regulator should be provided by customer.
Temperature 20 to 25℃(fluctuation: 1°C/hor less)
Humidity 60% or less
  • Please provide a table for mounting the equipment.

  • Specifications subject to change without prior notice.

Application

Application EM-09100IS

More Info

JEOL Instrument Basics

Easy explanation about mechanisms and
applications of JEOL products

Close
Notice

Are you a medical professional or personnel engaged in medical care?

Yes

No

Please be reminded that these pages are not intended to provide the general public with information about the products.

Contacts

JEOL provides a variety of support services to ensure that our customers can use our products with peace of mind.
Please feel free to contact us.