Close Btn

Select Your Regional site

Close

Analysis of volatile organic compounds inside a vehicle using thermal desorption-GC-MS [GC-QMS Application]

MSTips No. 435

Introduction

Thermal desorption (TD) is a GC pretreatment device for introducing gas samples such as atmospheric air and indoor air. It can also be used for gases generated from solid or liquid samples. Figure 1 shows an overview of TD operation. ① Sampling: Using a pump, etc., the gas sample is passed through and collected in a sample tube containing an adsorbent. ②Tube desorption: The sample tube is heated in the TD device, and the desorbed gas is collected in a cooled focusing trap. ③ Trap desorption: The focusing trap is heated at high speed and the desorbed gas is introduced into the GC in a narrow band width. These two steps of thermal desorption make it possible to obtain high sensitivity and sharp chromatogram peaks.

In this MSTips, we introduce the analysis of volatile organic compounds (VOCs) inside a vehicle, which is one of the main applications of TD-GC-MS. TD-100xr (Markes International Ltd) and JMS-Q1600GC (JEOL) were used in the experiment. The analysis conditions were based on the international standards ISO 12219-1 and JASO Z125.

 

Figure 1

Figure 1 Overview of TD operation

Features of TD-100xr

  • DiffLok cap: No need to remove from sample tube during measurement

  • Electronically cooled focusing trap: Cools down to -30 °C without refrigerant

  • Sample recollection: Collect the split flow during trap desorption and reuse the sample

Figure 2 TD-100xr
Figure 2 Focusing trap
Figure 2

Figure 2 Features of TD-100xr

Target component

We measured 7 components of VOCs which have guideline values of concentration inside a vehicle and  are specified TD-GC-MS as the measurement method. We also measured three components which are under additional consideration. Table 1 shows the guideline values of concentration established by the Ministry of Health, Labor and Welfare.

Table 1 Guideline values of concentration ( Left : current 7 components, Right : additional 3 components)

Name Concentration (μg/m3)
Toluene 260
Xylene 200
Ethylbenzene 3800
Styrene 220
Tetradecane 330
Dibutyl phthalate : DBP 17
Bis (2-ethylhexyl) phthalate) : DEHP 100
Name Concentration (μg/m3)
2-Ethyl-1-hexanol 130
Texanol 240
2,2,4-Trimethyl-1,3-pentanediol
diisobutyrate : TXIB
100

Experiment

For the current 7 components, 1, 10, and 100 ng of mixed standard solutions were added to sample tubes. For the additional 3 components, 50 ng of the mixed standard solution was added to the sample tube. The sampling volume of 3L described in ISO12219-1 corresponds to concentrations of 0.3, 3, 30 μg/m3 and 15 μg/m3. Additionally, 47 ng of Toluene-d8 was added to the sample tube as an internal standard. By using a special tool when adding these standard solutions, it is possible to easily collect the entire amount (Figure 3). Table 2 shows the measurement conditions of TD-GC-MS. In this measurement,  we measured the current 7 components and additional 3 components separately, but these can also be measured simultaneously.

Figure 3

Figure 3 Solution loading tool

Table 2 Measurement conditions

TD conditions
Thermal Desorption TD-100xr
(Markes International Ltd)
Sample tube type Tenax TA
Tube desorption 280°C (10 min), 30 mL/min, Splitless
Focusing trap type General purpose Hydrophobic (T2)
Trap cooling 0°C
Trap desorption 280°C (3 min), 2 0mL/min, Split 20:1
Flow path temperature 200°C
GC conditions
Gas Chromatograph 8890A GC
(Agilent Technologies)
Column ZB-5MSi (Phenomenex)
30 m x 0.25 mm, 0.25 μm
Oven Temperature 40°C (2 min)-15°C/min
-320°C (10 min)
Carrier flow He, 1.0 mL/min
MS conditions
Spectrometer JMS-Q1600GC (JEOL Ltd.)
Ionization EI+:70 eV, 50 μA
Ion source temperature 250°C
Mass range Scan, m/z 45-500

Measurement results of current 7 components

Figure 4 shows the TIC chromatogram at 100 ng/tube. Since xylene has 3 isomers as 2 peaks, the total number of peaks is 8.

 

Figure 4

Figure 4 TIC chromatogram of 100 ng/tube

 

Figure 5 shows the extracted ion chromatograms (EICs) and calibration curves of DBP (m/z 149) and DEHP (m/z 149). Good peak shapes and S/N sensitivity were obtained for all components at the lowest concentration. The determination coefficients of the calibration curve were 0.999 or higher, indicating good linearity.

 

Figure 5

Figure 5 EICs chromatograms and calibration curves (Left : DBP, Right : DEHP)

Measurement results of additional 3 components

Figure 6 shows the TIC chromatogram of 50 ng/tube. Texanol has two isomers as two peaks, the total number of peaks is four.

 

Figure 6

Figure 6 TIC chromatogram of 50 ng/tube

 

Figure 7 shows the extracted ion chromatograms (EICs) and mass spectra of each component. Good peak shapes and S/N sensitivity were obtained for all components. *Actually, Texanol-1 and Texanol-2 have different structures, but only one structure was registered in the library database, and both peaks matched that structure.

 

Figure 7

Figure 7 EICs and mass spectra

Conclusion

Using the TD-100xr and JMS-Q1600GC, it is possible to analyze the current 7 components and additional 3 components of VOCs inside a vehicle with high sensitivity and linearity. These devices are expected to be useful in analyzing gas samples such as atmospheric air and indoor air.

 

Solutions by field

Related information

JMS-Q1600GC UltraQuad™ SQ-Zeta Gas Chromatograph Quadrupole Mass Spectrometer

Close
Notice

Are you a medical professional or personnel engaged in medical care?

Yes

No

Please be reminded that these pages are not intended to provide the general public with information about the products.

Science Basics

Easy explanation about mechanisms and
applications of JEOL products

Contacts

JEOL provides a variety of support services to ensure that our customers can use our products with peace of mind.
Please feel free to contact us.