Close Btn

Select Your Regional site

Close

Rapid analysis of phthalates for RoHs2 by Py-GC-MS

MSTips No. 298

MSTips No. 298

Overview

Phthalates are used as plasticizers for resins, but their use is restricted by various regulations due to concerns about their effects on the human body as endocrine disruptors. The revised RoHS Directive (RoHS2), which came into effect on July 22, 2019, regulates four phthalats (diisobutyl phthalate: DIBP, dibutyl phthalate: DBP, butyl benzyl phthalate: BBP, di-2-ethylhexyl (DIBP, DBP, butyl benzyl phthalate: BBP, di-2-ethylhexyl phthalate: DEHP) were regulated. As measurement methods for these substances, the solvent extraction-GC/MS method for precise analysis and the Py-GC/MS method for screening are specified in IEC-62321-8. We have attempted a rapid analysis of Py-GC/MS method using JMS-Q1500GC by optimizing Py and GC oven conditions, and achieved a cycle time of 14.75 min per measurement. Here, we report the results.

Figure1.  Cycle time details

Figure1. Cycle time details

Experiment

PVC (NMIJ CRM 8152-a, about 0.1%), a certified reference material for the analysis of phthalates, was used as the measurement sample. A 0.50 mg sample was measured by Py-GC/MS to confirm the peak separation of the 4 phthalates and the repeatability of the area values at n=5. Table 1 shows the measurement conditions for Py-GC/MS.

Table 1. PY-GC/MS measurement conditions

PY
Sample weight 0.50mg
Furnace Temp. 150°C→200°C/min→320°C(0.5min)
GC
Column ZB-1HT 15m length, 0.25mmi.d.
0.1μm film thickness
(Phenomenex Inc.)
Gas flow He 1.0mL/min, constant flow
Injection mode Split 1/50
Inlet Temp. 300°C
Oven temp. 80°C→50°C/min→200°C→25°C/min→350°C
MS:JMS-Q1500GC
Ion source temp. 250°C
Interface temp. 300°C
Ionization EI, 70eV
Measurement mode Scan m/z 50 to 600 with SIM
 DIBP: m/z 149, 205, 223
 DBP: m/z 149, 205, 223
 BBP: m/z 91, 149, 206
 DEHP: m/z 149, 167, 279

※Under line : Quantitative ion

Results

The TIC chromatogram is shown in Figure 2. The 4 phthalates could be well separated.

Figure2.  TIC Chromatogram

Figure2. TIC Chromatogram

Next, the extracted ion chromatogram (EIC) of the quantitation ion for each of the 4 phthalates is shown in Figure 3. All components showed good detection sensitivity.

Figure3. Extracted Ion Chromatograms of phthalates in PVC

Figure3. Extracted Ion Chromatograms of phthalates in PVC

Finally, the area values of each measured sample at n=5 are shown in Table 2. For each phthalate, good results were obtained with a coefficient of variation of less than 10%.

Table 2.  Repeatability of area value

  #1 #2 #3 #4 #5 Average Standard deviation CV%
DIBP 178051 168249 158209 157911 167761 166036 8359 5.0%
DBP 166170 150082 152769 142665 148792 152096 8695 5.7%
BBP 210573 212346 205639 205776 243147 215496 15735 7.3%
DEHP 110307 108873 97884 93277 102132 102495 7212 7.0%

Summary

A rapid analysis of four phthalates, which are regulated by RoHS2, by Py-GC/MS method was attempted and the data was obtained with a cycle time of 14.75 min based on good detection sensitivity and repeatability.

Solutions by field

Related products

JMS-Q1600GC

JMS-Q1600GC UltraQuad™ SQ-Zeta Gas Chromatograph Quadrupole Mass Spectrometer

Close
Notice

Are you a medical professional or personnel engaged in medical care?

Yes

No

Please be reminded that these pages are not intended to provide the general public with information about the products.

JEOL Instrument Basics

Easy explanation about mechanisms and
applications of JEOL products

Contacts

JEOL provides a variety of support services to ensure that our customers can use our products with peace of mind.
Please feel free to contact us.