Delta Tips: Filename (Part 1)

NMDT_0003

1. How to customize filenames

JEOL NMR data are automatically saved in the JEOL data format (∗.jdf) under the filename defined in the storage_filename input box in the Header tab. This input box can contain very useful strings. For example, if the $(SAMPLE)_$(EXP.filename) string is used, the storage_filename is ‘sample-name_experiment-name’. In the figure below, the sample name is ‘CAHE’ while the experiment to be run is ‘proton’, and hence the storage_filename is ‘CAHE_proton’.

Delta Tips
  • The storage_filename can also contain the date of experiment if you use this string:
    $(YEAR)$(MONTH_NUM)$(DAY_NUM)_$(SAMPLE)_$(EXP.filename)
  • Default value in the storage_filename input box can be set in Config - Spectrometer Preferences - Environment – Filename Pattern. See ③ for details. This method does not apply to automation.

2. How to save data files in particular folder

You can have your data files saved in a specified subfolder inside the Data folder. To do so, type the subfolder name into the Project input box of Job Parameters (see the figure below).

Delta Tips

3. How to automatically include date in filename

First of all, log in as the Administrator of the instrument.
① Select Config - Instrument Preferences in the Spectrometer Control window.

Delta Tips

② Select the Environment tab. The parameter to modify is Filename Pattern.
If you input the $(YEAR)$(MONTH_NUM)$(DAY_NUM)_$(SAMPLE)_$(EXP.filename) string, the default storage_filename would be ‘YYYYMMDD_sample-name_experiment-name’ (for example, ‘20160623_CAHE_proton’).

Delta Tips

③ Select Parameters – Close to close the Instrument Preferences window.

Delta Tips
! The method shown above does not apply to data files to be collected by the use of automation templates. The method only applies to data files to be collected in the manual mode.
If you want to see a printed version,
please click this PDF file.

PDF 600KB
Close
Notice

Are you a medical professional or personnel engaged in medical care?

Yes

No

Please be reminded that these pages are not intended to provide the general public with information about the products.

JEOL Instrument Basics

Easy explanation about mechanisms and
applications of JEOL products

Contacts

JEOL provides a variety of support services to ensure that our customers can use our products with peace of mind.
Please feel free to contact us.