How to Set the Recycle Delay to Maximize the Sensitivity per Hour: Most Efficient Measurements

NM130016E

What should we do to obtain the maximum signal sensitivity in a given measurement time ? How about setting the recycle delay RD to be 3-5 times of the longitudinal relaxation time T1 ? It is the case of quantitative NMR measuements. For the pulse sequences preparing transverse magnetizations, the S/N ratio per hour is given by the following equation:

パルスシーケンス

The behavior of the function shown left clarifies that the signals can be most efficiently obtained when the recycle delay RD is set to be about 1.3 times of T1. In contrast, too long RD decreases the number of scans per hour and so lowers the sensitivity. Also for too short RD, the sensitivity per hour becomes low since the magnetizations are not sufficiently recovered.

Figures shown below are a series of 13C CPMAS spectra for sucrose (T1(1H) = 50s) obtained in a given time of 2,400s. The most efficient measurement is realized for RD=60s.

スクロース(T1(1H) = 50s)に対して ⼀定の測定時間2,400sで測定した13C CPMAS  スペクトル

References

  1. T. D. W. Claridge, “High-Resolution NMR Techniques in Organic Chemistry”, Second Ed., Elsevier, 2009, pp99-102.
  2. T. Nakai, New Glass, 28(2), 17-28 (2013).

SEARCH APPLICATIONS

Close
Notice

Are you a medical professional or personnel engaged in medical care?

Yes

No

Please be reminded that these pages are not intended to provide the general public with information about the products.

JEOL Instrument Basics

Easy explanation about mechanisms and
applications of JEOL products

Contacts

JEOL provides a variety of support services to ensure that our customers can use our products with peace of mind.
Please feel free to contact us.