Long-time VT attachment Compressor-type low-temperature unit

NM090006E

Product image※1

Features

・Easy operation for low temperature.
・Maintenance free (no movable parts).
・Supply ca. 25℃ colder air than RT to a probe.※2
・Long-time stability of temperature.
・No need for cooling media such as liquid nitrogen.

Installation View

Working principle

A compressor-type low-temperature unit generates cold air using Ranque-Hilsch vortex tube, which was discovered by Georges Ranque in 1930’s and became widely known in 1940’s by Rudolph Hirsch’s research. Nowadays, vortex tubes are mainly used to cool products and industrial tools in the area of machining. It generates air flow shown in Fig.1, yielding hot air and cold air. Its application to NMR was reported by R. W. Martin and K. W. Zilm in 2004 [1].

Fig.1 Air flow in vortex tube

Usage example 1

A compressor-type low-temperature unit can stabilize temperature, similarly to JEOL long-time VT equipments. As an example, Fig.2 shows NMR results of methanol under temperature control at 10℃ for 12 hours using JNM-ECA500. The peak positions of methyl and hydroxyl signals did not change, demonstrating high stability of temperature.

Fig.2 Time dependence of chemical shifts of methanol

Usage example 2

Figure 3 shows results of variable temperature NMR measurements, illustrating the chemical-shift difference of methyl and hydroxyl peaks of methanol at various temperatures. The linearity of the chemical-shift difference and temperature manifests applicability of the compressor-type low-temperature unit to variable temperature NMR measurements with high stability.

Fig.3 Results of variable temperature NMR measurements

※1:Specification and appearance may change without notice.
※2:Controllable temperature range depends on the instruments in use and environments.

Reference

[1] R.W. Martin, K.W. Zilm, Variable temperature system using vortex tube cooling and fiber optic temperature measurement for low
temperature magic angle spinning NMR, J. Magn. Reson. 168 (2004) 202–209.

SEARCH APPLICATIONS

Solutions by field

Close
Notice

Are you a medical professional or personnel engaged in medical care?

Yes

No

Please be reminded that these pages are not intended to provide the general public with information about the products.

JEOL Instrument Basics

Easy explanation about mechanisms and
applications of JEOL products

Contacts

JEOL provides a variety of support services to ensure that our customers can use our products with peace of mind.
Please feel free to contact us.