Strong interaction between light and electrons (4) "Transmission ESR/FMR measurement method"

ER200009E

Interaction between microwave photon and spins means in other words, interaction between resonating photon in a closing space and spins. Thus, measuring spectra using a non-resonant device can prevent the Purcell effect or strong coupling from distortion of intrinsic spectra. Figure 1(a) is a drawing expressing a typical spectroscopy. A light source irradiates light to a sample, and transmitted light is detected. A simple and non-resonant wave guide, as shown in Fig.1(b), can provide absorption spectra of para- and ferromagnetic samples which have a high spin density.

Fig.1 (a) A drawing of transmit absorption measurement. (b) An example of transmission ESR/FMR measuring device using a wave guide.

Fig.1 (a) A drawing of transmit absorption measurement. (b) An example of transmission ESR/FMR measuring device using a wave guide.

Transmission FMR spectrum

Spectra in Fig.2 are an FMR spectra of YIG-thin film (20μm thickness, conf. Application Note ER200008E ) using a detection device as shown in Fig.1(b). Merits of this method are that tuning of the filling factor and magnetic modulation are not necessary. Furthermore, dependence of FMR spectra on the frequencies is easily obtained. Application Note ER200010E will explain it more.

Fig.2 An example of transmission FMR spectra using YIG thin film.

Fig.2 An example of transmission FMR spectra using YIG thin film.

Please see the PDF file for the additional information.
Another window opens when you click.

PDF 404.0KB

SEARCH APPLICATIONS

Close
Notice

Are you a medical professional or personnel engaged in medical care?

Yes

No

Please be reminded that these pages are not intended to provide the general public with information about the products.

JEOL Instrument Basics

Easy explanation about mechanisms and
applications of JEOL products

Contacts

JEOL provides a variety of support services to ensure that our customers can use our products with peace of mind.
Please feel free to contact us.