Close Btn

Select Your Regional site

Close

Very Low Temperature Measurement - Liquid He Unit ES-CT470

ER080003E

One of the features of ESR spectroscopy is that the lower the sample temperature, the higher the sensitivity becomes. This is because the number of spins in the two spin states, split by Zeeman interaction, becomes greater the lower the temperature. In some samples e.g. some metal complexes, the spin-spin relaxation time (T2) is too short to measure at room temperature and require very low temperature to measure the ESR spectrum. In such cases, measurement at low temperature is useful and the ES-CT470 liquid He VT unit is a useful attachment for such measurements.

The photo shows the ES-CT470 set-up. The main features of the system are:-

  • 1. Easy set up.
        Installation time ~ 20 min.
  • 2. Fast cool down
       Room temp to 4.5 K in ~ 15min.
  • 3. High temperature stability
       士0. 0l K (2.5 to 4.2 K)
       士0.5K (4.2 to 470 K)
  • 4. Liquid He supply from rear of magnet
       Good front access.
       Optical-irradiation possible from front of magnet
  • 5. Easy sample exchange
the ES-CT470 set-up
ES-CT 470 Setup Dewar is not included in the basic supply

The figure shows the temperature dependence of Fe3+  signal in glass. For paramagnetic species, the signal intensity is inversely proportional  to the absolute temperature (Curie’s law.) For samples such as natural products which can be obtained only in very small quantities, high sensitivity by lowering temperature is clearly advantageous.

Temperature Dependence of Fe3+ Signal
Temperature Dependence of Fe3+ Signal

SEARCH APPLICATIONS

Close
Notice

Are you a medical professional or personnel engaged in medical care?

Yes

No

Please be reminded that these pages are not intended to provide the general public with information about the products.

JEOL Instrument Basics

Easy explanation about mechanisms and
applications of JEOL products

Contacts

JEOL provides a variety of support services to ensure that our customers can use our products with peace of mind.
Please feel free to contact us.