High resolution TEM image by JEM-1400Flash

EM2020-11E

The optical system of the JEM-1400Flash can image at an ultra-high resolution of 0.14 nm while having a magnification range (minimum magnification: 10 times) that can image the entire sample mesh (φ2 mm) with a bottom mount camera.
The objective lens is de-excited at an extremely low magnification so that the entire sample mesh can be observed. On the other hand, at high magnification, the objective lens is excited and the image plane created below the objective lens is magnified by the imaging lens, so that a maximum magnification of 1.5 million times can be obtained (Fig. 1).
At this time, since the objective aperture is inserted in the back focal plane of the objective lens, a high-contrast image can be obtained.
In order to obtain a low-magnification image of several hundred times with the objective lens excited, the objective mini (OM) lens is excited and the image plane is reduced (Fig. 2). In this case as well, the objective aperture that determines the contrast is inserted in the back focal plane, so the same contrast can be obtained from low to high magnification. The JEM-1400Flash can change the magnification smoothly from extremely low magnification to a maximum magnification of 1.5 million times without image rotation, and it is always possible to observe with high contrast.
A high-resolution TEM image taken with JEM-1400Flash, which has an optical system that can easily obtain low-magnification images and high-magnification images, is shown (Fig. 3).

Fig. 1 When exciting the objective lens to create a high magnification

Fig. 1 When exciting the objective lens to create a high magnification

Fig. 2 When exciting an OM lens to create a low magnification

Fig. 2 When exciting an OM lens to create a low magnification

図3. JEM-1400Flash (HR) で撮影した高分解能像 (試料:金単結晶、加速電圧:120 kV)
Sample : Au single crystal
Accelerating voltage: 120 kV
図3. JEM-1400Flash (HR) で撮影した高分解能像 (試料:カーボンナノチューブ、加速電圧:120 kV)
Sample : Single layer carbon nanotube
Accelerating voltage: 120 kV
図3. JEM-1400Flash (HR) で撮影した高分解能像 (試料:グラファイト、加速電圧:120 kV)
Sample : Graphite
Accelerating voltage: 120 kV

Fig. 3 High-resolution image taken with JEM-1400Flash (HR)

Please see the PDF file for the additional information.
Another window opens when you click.

PDF 2.43MB

Solutions by field

Close
Notice

Are you a medical professional or personnel engaged in medical care?

Yes

No

Please be reminded that these pages are not intended to provide the general public with information about the products.

JEOL Instrument Basics

Easy explanation about mechanisms and
applications of JEOL products

Contacts

JEOL provides a variety of support services to ensure that our customers can use our products with peace of mind.
Please feel free to contact us.