【DISCONTINUED】JIB-4601F MultiBeam System

DISCONTINUED

JIB-4601F MultiBeam System

This product is no longer available.

If you would like to know the latest information about your preferred product or to find out more about alternatives, please click on the link below. We hope you will continue to use our products.

The JIB-4601F is a MultiBeam processing and observation system that incorporates a Schottky-type SEM and a High-Power FIB column. After performing cross-section milling of a region using FIB, it is possible to observe and analyze the internal portions using SEM. The FIB can be used for fine milling and TEM sample preparation, as well as performing 3-dimensional structure analysis with automatic cross-section milling/ observation & analysis at fixed intervals. The Schottky-type SEM enables easy positioning of the target site, as well as observation and analysis of minute structures. This is a system that can be used for a wide range of applications, including research & development as well as product quality control.

Features

All-in-one tool – Various applications offered by one instrument

Multi-function analysis through the combination of a High-Power FIB column and High-Power Optics SEM

The High-Power FIB column, with a maximum probe current of 60 nA, enables fast cross-section milling, while the Schottky SEM offers high-resolution observation of the cross-section. The High-Power Optics (maximum current 200 nA @ 15 kV) that are incorporated in the SEM make it possible to perform high-speed, high-resolution analyses, such as element analysis using EDS or WDS, and EBSD crystal orientation analysis. In addition, there is a full range of optional attachments available, including transmission electron detectors, specimen cooling mechanisms like a cryo unit and cool stage, and cathodoluminescence detectors.

3-dimensional structure analysis function

By automatically alternating between FIB section milling using the serial slicing and sampling function, and SEM observation and analysis (EDS mapping or EBSD crystalline structure analysis) it is possible to analyze the 3-dimensional structure of a specimen. The measurements can be performed rapidly because there is no need to shift the specimen stage.

Compatible with other JEOL products

The specimen holders used with the JIB-4601F are the same as those used by the JEOL FE-SEM systems. Furthermore, if a shuttle retainer is used between the TEM holder tip and the JIB series stage, there is no need for complicated handling of a TEM sample that is mounted to the mesh grid, offering improved throughput.

Specifications

Electron gun Schottky Field Emission Type △TFE(ZrO/W)
Objective lens Out-lens type objective lens
Resolution 1.2 nm @ 30 kV WD4mm
3.0 nm @ 1 kV WD2mm
Accelerting Voltage 0.2 to  30kV
Probe current Several pA to 200nA @ 15 kV
Aperture angle optimizing lens Built-in
Magnification x20 to x1,000,000
Detectors Lower Electron Detector(LED)
Option(*)
R-BED*
STEM*
Gentle beam Built-in (Voltage applied to specimen  0 to 300V)
Specimen exchange chamber Built-in
(with dry nitrogen introduction function)
Specimen stage 6-axis motor drive stage
Specimen travel range X: 50 mm , Y: 50 mm,
Z:  1.5 to 41 mm
Tilt: -5 to 70°, Roration: 360°,
Fz: -3 to +3mm
Specimen exchange method One-touch chucking type
Specimen holder Standard Holder :
φ12.5 holder
φ32mm holder
Option holder :
76.2mm wafer holder
 100mm wafer holder
 125mm wafer holder
 150mm wafer holder
2,4,6 type bulk holder
Surface multi-specimen holder
STEM holder
Shuttle retainer holder
One-touch holder
Chamber scope Specimen Chamber Camera (option)
Auto functions Auto focus
Auto brightness control
System control SEM controlling system
   PC IBM PC/AT compatible〔SM-77360PC〕
RAM 2 GB or more
 OS Windows 7® Professional*
Monitor 23 inch
Image display Image display area 1280×960 pixel, 800×600 pixel
Display modes Standard : SEM_SEI, FIB_SEI
Option : SEM_COMPO, SEM_TOPO, AUX, CCD
Scanning / Display modes Limited area scanning, addition iamge, scaler
Screen display: 1 screen, 2 screen(standard), 4 screen
Vacuum HV ; < 2.0 x 10E-4 Pa
(when GIS is in use;<3.0E-3Pa) 
Evacuation system SIP x 2 (SEM), 
SIP x 1 (FIB),
TMP x 1, v RP x 1
Electricity consumption saving mode In normal operation approx. 2.0kVA
Co2emission equivalent         Co2 emission in year
Normal operation       4967kg
Safety device Protection against vacuum fall, water failure, N2 gas pressure fall, leakage current
However, ultra high vacuum maintenance mechanism at power failure is not included.
Footprint 3200mm or more ×3000mm or more
Installation Requirement Power Supply Single PHase 100 V, 50/60 Hz, max. 4kVA,
Normal use approx. 2.0kVA
Allowable input power supply fluctuation within ±10%
Grounding terminal  100Ω or less  x 1
Cooling water
Faucet out. dia. 14 mm x 1, or JIS B 0203 Rc 1/4  x 1
Flow rate 0.5 L/min
Water pressure 0.1to 0.25 MPa (gauge pressure) Water temperature 20°± 5℃
Drain inn. dia.  25 mm  or more x1, or JIS B 0203 Rc 1/4  x 1
Dry nitrogen gas JIS B 0203 Rc 1/4(to be provided by user)
 Pressure 0.45 to 0.55 MPa (gauge pressure) 
Installation Room 
 Room temperature 20℃±5℃
 Humidity 60%  or lower
 Stray magnetic field 0.3μT (P-P) or less  (50/60 Hz sin wave,WD 10 mm,15 kV)*1
 Floor vibration 3μm (P-P) or lower*1 at frequency of sin wave more than 5Hz.
 Noise 70 dB or lower *1 as FLAT characteristic
 Dimension of installation room 3,000 mm x 3,000 mm or more
Height 2,300 mm or higher
 Door size 1,000*2 mm(w) x 2,000 mm(H) or more
  • This is not applicable for conditions other than these.  We will conduct installation room survey before delivery, and consult with you the maximum magnification of observation. 

  • Please consult if door width is 900 mm.

Main optional attachments EDS, WDS, EBSD, CL, SNS  etc.
  • Windows is a registered trademark of Microsoft Corporation in the United States and other countries.

More Info

JEOL Instrument Basics

Easy explanation about mechanisms and
applications of JEOL products

Close
Notice

Are you a medical professional or personnel engaged in medical care?

Yes

No

Please be reminded that these pages are not intended to provide the general public with information about the products.

Contacts

JEOL provides a variety of support services to ensure that our customers can use our products with peace of mind.
Please feel free to contact us.