Observation Examples of Processed Foods

SM064

Foods observed with SEM

Specimen deformation during SEM observation of food, due to the fat and water content, can occur either by vacuum conditions in the SEM specimen chamber or by thermal effects caused by electron beam irradiation. In order to counteract these effects, the cooling of the specimen before SEM observation is required with a cryo stage or LV cooling holder.
The cryo stage can cool and maintain the specimen at around liquid nitrogen temperature, allowing for observation and analysis of the specimens containing a lot of fats or water (mayonnaise, fresh cream, etc.). However, a conductive coating is often required.
The LV cooling holder allows for precooling the specimen with liquid nitrogen, although does not maintain this temperature. However, the LV cooling holder can be easily combined with low vacuum SEM, allowing for observation and analysis of uncoated samples with a lower water or fat content (raw meats, vegetables, etc.) for a short period of time.
This report presents morphological observation examples of some processed foods, with a cryo stage or LV cooling holder under low vacuum. Specimens used were carbohydrates, frozen foods and meat products.

Carbohydrates
Fractured surface of boiled noodles
Cooled with LV cooling holder

Frozen foods
Fractured surface of cream in croquette
Cooled with Cryo stage

Meat products
Fractured surface of chicken
Cooled with Cryo stage

Cross section of baked bread
Cooled with LV cooling holder

Fractured surface of whipped cream
Cooled with Cryo stage (Ice cross section)

Fractured surface of ham
Cooled with Cryo stage

Please see the PDF file for the additional information.
Another window opens when you click.

PDF 998KB

Solutions by field

Close
Notice

Are you a medical professional or personnel engaged in medical care?

Yes

No

Please be reminded that these pages are not intended to provide the general public with information about the products.

JEOL Instrument Basics

Easy explanation about mechanisms and
applications of JEOL products

Contacts

JEOL provides a variety of support services to ensure that our customers can use our products with peace of mind.
Please feel free to contact us.