Close Btn

Select Your Regional site

Close

High sensitivity and quantitative 13C measurements using "Q-POMMIE"

NM190016E

POMMIE (Phase Oscillations to MaxiMIze Editing) is a 13C experiment that, like the more familiar DEPT experiment, utilizes polarization transfer to enhance the intensities of the 13C signals. However, unlike DEPT, POMMIE edits the spectrum by varying pulse phase rather than adjusting pulse flip angle.
Fig. 1 shows the pulse program of Q-POMMIE (Quantitative-POMMIE)1). It improves quantitative performance by varying Δ and pulse phase to average the efficiency of polarization transfer (Fig. 1). The spectrum pattern of Q-POMMIE is identical to DEPT45.

Quantitative NMR measurements require sufficiently long repetition times to allow the (near complete) recovery of magnetization between scans. For the standard 13C{1H} inverse gated decoupling method, the minimum acceptable repetition time is dictated by the 13C T1s, which can be very long (up to several minutes). On the other hand, the minimum repetition time of the Q-POMMIE method is dictated by the usually much shorter 1H T1s. This means that Q-POMMIE can yield quantitative 13C spectra in less time and with higher sensitivity than via the 13C{1H} inverse gated decoupling method.

Fig. 2 shows a comparison of 13C{1H} inverse gated decoupling and Q-POMMIE spectra of 10% Cinnamic acid cis-3-hexenylester in CDCl3. Although quaternary carbons are not observable in the Q-POMMIE spectrum, protonated carbon signals are recorded with significantly higher sensitivity, thereby allowing more accurate quantitation. Tab. 1 shows the comparison of the peak integrals obtained from each spectrum. While the CH integrals are significantly underrepresented in DEPT45, those obtained using Q-POMMIE are very close to those obtained via the 13C{1H} inverse gated decoupling method.

Features of Q-POMMIE
(Comparison with 13C{1H} inverse gated decoupling)

  • Higher sensitivity
  • Quaternary carbons are not observed
  • Quantitative condition is dictated by 1H T1s not 13C T1s
  • Requires large minimum number of scans (96×n)
Fig. 1: Pulse program of Q-POMMIE

Fig. 1: Pulse program of Q-POMMIE

Fig.2:  a ) 13C{1H} inverse gated decoupling

Fig.2: a) 13C{1H} inverse gated decoupling

b) : Q-POMMIE

b) : Q-POMMIE

Sample:10% CAHE/CDCl3
Instrument:JNM-ECZ400S & ROYALprobe™HFX
Number of scans:384
Pulse repetition time:46s

Tab.1: Integral values of protonated 13C signals shown in Fig. 2

δ(ppm)/Exp 13C{1H} inverse gated decoupling Q-POMMIE DEPT45
14.3(CH3) 1.00 1.00 1.00
20.7(CH2) 0.98 0.97 0.92
26.9(CH2) 0.89 0.94 0.94
64.2(CH2) 0.99 1.07 1.07
118.3(CH) 0.95 0.97 0.69
123.9(CH) 0.91 0.92 0.67
128.1(CH)*2 1.85 1.87 1.44
129.0(CH)*2 1.96 1.92 1.45
130.3(CH) 0.95 0.99 0.67
134.5(CH) 0.87 0.85 0.64
144.8(CH) 0.95 0.96 0.66

Practical Example: UV Initiator

As an illustration of the utility of the Q-POMMIE experiment, Fig. 3 shows a comparison of spectra recorded on a sample of a UV initiator. The 13C spectrum of this sample contains a signal at 77.2 ppm which is completely obscured by the solvent signal, so its integral cannot be determined via 13C{1H} inverse gated decoupling (Fig. 3a). On the other hand, the solvent signal is not visible in the Q-POMMIE spectrum, allowing this signal to be cleanly observed and hence integrated (Fig. 3b).

Fig. 3: a) <sup>13</sup>C{<sup>1</sup>H} inverse gated decoupling b) Q-POMMIE

Fig. 3: a) 13C{1H} inverse gated decoupling b) Q-POMMIE

Sample preparation: 10mg sample/CDCl3
Instrument: JNM-ECZ500R & 5mm SuperCOOL probe
Scans: 960
Pulse repetition time: 20s

Reference

1) Anal. Chem. 2008, 80, 8293-8298.

Sample courtesy of Mr. Yuuji Itoh (TOYO INK SC HOLDINGS CO., LTD)

Please see the PDF file for the additional information.
Another window opens when you click.

PDF 3.84MB

SEARCH APPLICATIONS

Solutions by field

Close
Notice

Are you a medical professional or personnel engaged in medical care?

Yes

No

Please be reminded that these pages are not intended to provide the general public with information about the products.

JEOL Instrument Basics

Easy explanation about mechanisms and
applications of JEOL products

Contacts

JEOL provides a variety of support services to ensure that our customers can use our products with peace of mind.
Please feel free to contact us.