Analysis of LIB cathodes by using MATPASS with ultra-fast MAS

NM170022

NMR is extremely useful for characterizing local structure of materials for lithium ion battery (LIB) because it is able to observe lithium ions directly . However, a solid state NMR analysis of cathode materials using 7Li by conventional MAS frequency (- 20 kHz) has been difficult due to a series of spinning side band (SSB) originated from paramagnetic shift anisotropy.
The ultra-fast MAS probes (1mm and 0.75mm) give high resolution and less SSB 7Li spectra as shown in an application note NM-140020. However, for many of LIB cathode materials the ultra fast MAS alone may not be sufficient to obtain spectra without effect of SSBs. An excellent solution to avoid the problem is MATPASS (Magic Angle Turning – Phase Adjusted Spinning Sideband) technique developed by Hung et al. 1 The experiment allows SSBs free 7Li spectra of paramagnetic cathode materials. This technique is available for moderate speed MAS probes, but the ultra fast MAS probe still has great advantage because it allows wide range excitation leading to quantitative spectrum and experiment with small sample amount.

Ultra fast MAS + MATPASS

  • SSB free 7Li spectrum
  • Wide range excitation
  • Small sample amount

1mm sample tube

1. I. Hung et al., J. Am. Chem.Soc. 2012, 134, 1898−1901

Deterioration analysis of LiMn2O4 by 7Li MATPASS spectra
LiMn2O4劣化前後の7Li MATPASSスペクトル
Samples before and after cycles were kindly provided by Prof. Y.Oka (Univ.Hyogo)
Please see the PDF file for the additional information.
Another window opens when you click.

PDF 814KB

SEARCH APPLICATIONS

Solutions by field

Close
Notice

Are you a medical professional or personnel engaged in medical care?

Yes

No

Please be reminded that these pages are not intended to provide the general public with information about the products.

JEOL Instrument Basics

Easy explanation about mechanisms and
applications of JEOL products

Contacts

JEOL provides a variety of support services to ensure that our customers can use our products with peace of mind.
Please feel free to contact us.