System Validation in Quantitative NMR

NM150006E

Factors contributing to reliability of analytical results

System Performance Affecting Quantitative NMR Data: S/N and Resolution

Fig. 1: S/N and repeated integral accuracy (SD)
Fig. 1: S/N and repeated integral accuracy (SD)

Fig. 1 shows the S/N and the accuracy of its integral value (SD: Standard deviation) that are theoretically calculated.The SD is the result of integration repeated 100 times. The figure shows that the integral accuracy is heavily dependent on the S/N; for example, an integral accuracy of 1% or less requires an S/N of 100 or more.

The S/N also affects resolution. Fig. 2 shows the results of qNMR experiments. The S/N of BTMSB, a standard for quantitative analysis, is reduced to 1/3 when resolution is low.

If signals having a low S/N are selected for quantitative analysis, resolution alone can significantly affect the results of analysis.

Fig. 2: S/N vs. resolution qNMR sample:Analyte: vinclozolin; quantitative reference: 1,4-BTMSBd6 deuterated DMSO solution
Fig. 2: S/N vs. resolution qNMR sample:Analyte: vinclozolin; quantitative reference: 1,4-BTMSBd6 deuterated DMSO solution

Quantitative NMR: Frequently asked questions

  • How accurate is quantitative data going to be in my NMR system?
  • Does my NMR system continue to maintain its performance level?
  • Why is the data different this time? Is my system operating at peak performance?

Daily System Checkup for Quantitative NMR Analysis

Operating procedure
  1. 1.Sample
    • The same sample is used for each experiment.
    • The analyte and reference are loaded.
    • The analyte and reference are reliable including SI traceability.
    • The sample is a stable solution.
  2. 2.Analytical process
    • The measurement method is validated.
  3. 3.Analytical results (according to the system compatibility guidance in Japanese Pharmacopoeia)
    • 3 samples: Repeat analysis 3 times or 6 times
    • Detection: S/N > 100
    • System compatibility: Molecular integral ratio < 1
    • System reproducibility: Relative standard deviation in area ratio < 1 %
Operating procedure

JEOL provides quantitative NMR menus for system diagnosis and periodical inspection.
Contact your local representative for details.

SEARCH APPLICATIONS

Close
Notice

Are you a medical professional or personnel engaged in medical care?

Yes

No

Please be reminded that these pages are not intended to provide the general public with information about the products.

Science Basics

Easy explanation about mechanisms and
applications of JEOL products

Contacts

JEOL provides a variety of support services to ensure that our customers can use our products with peace of mind.
Please feel free to contact us.