Long-term measurement using variable temperature controller (ES-13060DVT5)

ER210010E

Variable temperature controller (ES-13060DVT5)

The variable temperature controller can be combined with the cavity to enable ESR measurements in the temperature range of -170 °C to +200 °C. The temperature is controlled by heating liquid nitrogen (low temperature range) or air (high temperature range). The 10 L dewar is the standard, the 30 L dewar is available as an option, which can be used for long-term temperature control. The continuous operation time at low temperature is approximately 70 minutes at -160 °C and approximately 90 minutes at -150 °C with a 10 L dewar, but with a 30 L dewar, it can be operated for approximately 3 times longer at any temperature. With this controller, the ESR measurement can be performed while changing the temperature as well as the fixed temperature. If the measurement temperature is to be changed automatically and continuously, it is necessary to readjust the microwave conditions (frequency, coupling, etc.) for each temperature, but it is possible to have the ESR spectrometer do it automatically.

Application

Using the quartz sample extracted from the sediment irradiated with gamma rays, automatic continuous ESR measurement was performed while raising the temperature by 2 °C in the range of -150 to 50 °C. The measurement time, including readjustment of the microwave conditions at each temperature, was approximately 7 hours. With a 10 L dewar, liquid nitrogen was depleted in about 5 hours, so the measurement could not be continued.
At -150 °C, the Al center and the Ti-Li center were clearly observed. As the temperature increased, the relaxation time (T2) of these paramagnetic centers rapidly shortened, and the signal intensity decreased with the disappearance of fine spectral structure due to broadening the line width (Figure 1). It is almost no longer observable at 24 °C. Figure 2 shows contour plots of all ESR spectra (101 spectra) from -150 to 50 °C. This figure shows the temperature dependence of the line width and the intensity for each component. Using the variable temperature controller with a 30 L dewar, it is possible to measure ESR signals at low temperature for long-term.

Figure 1

Figure 1. Temperature dependence of quartz ESR signal.

Figure 2

Figure 2. Temperature dependence of quartz ESR signal.

Please see the PDF file for the additional information.
Another window opens when you click.

PDF 656.4 KB

SEARCH APPLICATIONS

Solutions by field

Close
Notice

Are you a medical professional or personnel engaged in medical care?

Yes

No

Please be reminded that these pages are not intended to provide the general public with information about the products.

JEOL Instrument Basics

Easy explanation about mechanisms and
applications of JEOL products

Contacts

JEOL provides a variety of support services to ensure that our customers can use our products with peace of mind.
Please feel free to contact us.