Paramagnetic vacancy of synthetic quartz glass

ER210009E

E' center signal

When the synthetic quartz sample tube is irradiated with ultraviolet rays, ESR signal due to defects is observed (Figure 1). This paramagnetic vacancy in the quartz glass is called the E1' center which has one electron at the center [1]. Several types of E' centers with different generation mechanisms have been reported so far. In addition, E1' center signal is observed by the mechanical stress as well. This signal is measured at low microwave power because of its long relaxation time. ESR signal intensity of this center varies depending on the raw material and the manufacturing process.

Figure1

Figure1. The E1' center signal of the quartz sample tube by ultraviolet irradiation.

Dependence of E1' center signal on intensity and time of UV irradiation

In Ultraviolet Radiation Apparatus (ES-13080UV2A / ES-13090UV04), UV irradiation intensity can be entered numerically. Figure 2 shows the time dependence of the E1' center signal intensity (A in Figure 1) when the sample tube is irradiated with different intensity by ES-13080UV2A light source. It can be seen that the longer the irradiation time and the higher the irradiation intensity, the higher the E1' center signal intensity. When performing ESR measurement while irradiating the sample with ultraviolet rays, it is necessary to confirm that the E1' center signal derived from the sample tube does not affect the ESR signal of the target paramagnetic species.

Figure 2

Figure 2. The E1' center signal intensity plotted against the irradiation time.

Refference

[1] R.Weeks, J. Non-Cryst. Solids,179,1 (1994).

Please see the PDF file for the additional information.
Another window opens when you click.

PDF 821KB

SEARCH APPLICATIONS

Solutions by field

Close
Notice

Are you a medical professional or personnel engaged in medical care?

Yes

No

Please be reminded that these pages are not intended to provide the general public with information about the products.

JEOL Instrument Basics

Easy explanation about mechanisms and
applications of JEOL products

Contacts

JEOL provides a variety of support services to ensure that our customers can use our products with peace of mind.
Please feel free to contact us.