Close Btn

Select Your Regional site

Close

Evaluation of Active Oxygen Scavenging Activity 1 O2・-

ER070002E

It is well known that some active oxygen species can cause oxidation reactions that are  problematic in biological systems, e.g. become a trigger for various diseases. Therefore, evaluation of active oxygen scavenging (antioxidation) activity is commonplace in fields such as drugs, food, supplement development, etc.
Here, we introduce a method to evaluate the scavenging activity of the superoxide anion radical ( O2・-) which is a common active oxygen species found in living bodies.

  • Reagent
    Reagent Maker Concentration Sampling Vol./Assay
    1.DMPO Labotec  Stock Solution 15μl
    2.Hyx Sigma -Aldrich 5 mM-H2O  50μl
    3.Buffer/DMSO(1/5dil) - 0.1 mol/l 35μl
    4.Standard (SOD)/sample a series of concentrations 50μl
    5.XOD Roche 0.4 U/ml-buffer 50μl
    Total 200μl
    DMPO      5,5-dimethyl-1-pyrroline-N-oxide (spin trap agent)
    Hyx          hypoxanthine,
    Buffer : 0.1 mole/l Phosphate buffer (pH 7.8)
    DMSO      to be added to reduce the influence of DMPO-OH co-existing on the spectrum.
    SOD         superoxide dismutase
    XOD        xanthine oxidase
  • Measurement Procedure:
    1. Make each of the above reagents in labelled tubes.
    2. Add reagents 1 to 4 and mix thoroughly for 1 to 2 seconds in a test tube, then add 5 and mix thoroughly for about 5 seconds.
    3. Place the solution into an aqueous cell and cap it, wipe the exterior, then insert it into the cavity.
    4. Start ESR measurement after a known time (e.g. 1 minute) after the addition of XOD to the solution.
    5. Measure the height of Mn2+ peak and DMPO-O2-peak from the spectrum, and calculate the relative intensity (R=DMPO-O2-/Mn2+) as shown in Fig.1.

    Fig.1 DMPO-O2- spectrum (when DMSO is added).
    Fig.1 DMPO-O2- spectrum (when DMSO is added).
  • ESR Measurement Conditions:
    Field: 335±5 mT;
    Power: 4 to 8 mW;
    Modulation Width: 0.079 mT;
    Sweep Time: 2 to 4 min;
    Time Const.: 0.1 to 0.3 sec; Amplitude: 250 to 400
  • DMPO-O2- Spectrum Analysis
    Fig. 2 shows the DMPO-O2- spectrum. Here, the unpaired electrons are localized on the N-O orbital and the signal is split into 12 lines due to hyperfine interaction with N, H-atom (β), and H-atom (γ). The weak signals observed on both sides of these signals are derived from co-existing DMPO-OH. In order to reduce the quantity of DMPO-OH, DMSO should be added.
    In this case, DMPO-CH3 (marked “+” in Fig.1, composed of 6 lines, 2 of which overlap with DMPO-O2- signals and are obscured) generated from the reaction of DMSO with HO・is observed.
    図2 DMPO-O2-スペクトルのバーグラム
    Fig.2 ESR spectrum of DMPO-O2-
  • Anti-oxidation Capability Evaluation
    When a sample with anti-oxidation activity is added to the reagent mixture series as above, DMPO-O2- signal will reduce according to the level of anti-oxidation activity. When the anti-oxidation activity of the sample is to be measured, create a calibration curve with a series of SOD concentrations. As SOD is a substance known to have a very strong anti-oxidant activity against O2・-, it is a good standard substance. Next, add the unknown sample solution to the solution in the same way and obtain R for each case, and then the anti-oxidation activity, as SOD-equivalent concentration, is obtained from the calibration curve. (Refer to ER040001 & ER040004).
  • Reagent Handling Precautions
    DMPO is not so stable and is decomposed by heat, ultraviolet, or oxygen to give background signals. As the melting point is 35°C only the amount required should be melted and the remainder refrozen. If the reagent bottle is then filled with nitrogen gas, it may be preserved for a relatively long time.
    As hyx is only sparingly soluble, it should be dissolved in water at 50°C. Preservation for about a month is possible if refrigerated.
    SOD is dissolved in phosphate-buffer.
    XOD, as the rot difference of O2・- generation activity is not negligible. As it is an unstable enzyme, it needs to be used within a month once opened.
    An example of SOD dilution series: 30, 15, 7.5, 3.75, 1.88, 0.94, 0 unit/ml.

SEARCH APPLICATIONS

Close
Notice

Are you a medical professional or personnel engaged in medical care?

Yes

No

Please be reminded that these pages are not intended to provide the general public with information about the products.

JEOL Instrument Basics

Easy explanation about mechanisms and
applications of JEOL products

Contacts

JEOL provides a variety of support services to ensure that our customers can use our products with peace of mind.
Please feel free to contact us.