Let’s Use ESR I - How to Select the Sample Tube

ER060004E

ESR may be used to measure a wide range of samples. However, it is necessary to determine the most appropriate conditions for each sample. The following measurement flow chart gives a step-by-step approach. Here, we explain how to select the most appropriate sample tube.

ESR Measurement Flow

ESR Measurement Flow

Sample Type and Sample Tube Selection

• Liquid, Solid, Gas・・・All are possible!
Straight Pipe Type Sample Tube of 4mm I.D.

Caution: Substance with Large Dielectric Loss
a.Aqueous Solution・・Flat Cell, Capillary Cell
b.Sample of high moisture content・・Dry, Use Tissue Cell
c.Conductive substances・・Decrease quantity, Use capillarγ

Fig.1 Straight Standard Sample Tube(100mm length tube is also available.)
 Straight Standard Sample Tube
The measurement section of the tube is made of high purity quartz. Tubes with the upper part of heat-resistant glass are suitable for glasswork.
Fig.2 Example of Sample Collection
Example of Sample Collection

(Left) Samples up to 43.5mm high can be measured.
(Middle) If there is more sample, there is no contribution to the signal.
(Right) If radical quantity is adequate, even a sample quantity of sample gives signal.

Flat Cell for Aqueous Solution Sample
Fig.3 Flat Cell for Aqueous Solution Sample
Tissue Cell for Tissue Sample
Fig.4 Tissue Cell for Tissue Sample

Solvents that have a large dielectric loss, e.g. water, chloroform, alcohols, sulfuric acid, etc., or samples that contain more than 10% of these solvents, cannot be measured using the standard sample tube. For a sample in solution, please use a flat cell (Fig.3), or if it is a solid sample, use thin pieces in a tissue cell. For measurement using a flat cell, about 200ul of solution is needed. For small amounts of sample, use a capillary tube inner diameter 0.8mm or less. Capillary tubes are also suitable as sample tubes for materials with a large dielectric loss. If the measurement is at room temperature, disposable tubes are available. But for low temperature measurements, a quartz capillary tube (2 mm, 2.6 mm , 3 mm O.D.) is suitable as shown in our parts catalogue.

SEARCH APPLICATIONS

Solutions by field

Close
Notice

Are you a medical professional or personnel engaged in medical care?

Yes

No

Please be reminded that these pages are not intended to provide the general public with information about the products.

JEOL Instrument Basics

Easy explanation about mechanisms and
applications of JEOL products

Contacts

JEOL provides a variety of support services to ensure that our customers can use our products with peace of mind.
Please feel free to contact us.