【DISCONTINUED】IB-09010CP Cross Section Polisher

DISCONTINUED

IB-09010CP Cross Section Polisher

This product is no longer available.

If you would like to know the latest information about your preferred product or to find out more about alternatives, please click on the link below. We hope you will continue to use our products.

Standard device for cross-section preparation using an ion beam, which is best suited to SEM observation and analysis of crystal orientation.

Features

The IB-09010CP can produce pristine cross sections of specimen that were difficult to handle using a mechanical grinding method. This device is especially suitable for the preparation of cross sections of complex materials in which hard substances and soft substances are intermingled.
The IB-09010CP is an Ion Milling device to perform cross section processing.
A function to automatically start the processing without any waiting time is included. In this way, the process can be started at the optimal vacuum rate.
There is an optional 8kV power supply available, which provides high processing speeds of 300μm/h (on a silicon substrate conversion basis, excrescence quantity : 100μm, 2 hours average)
Easy operation by touch panel

Specifications

Ion accelerating voltage 2 to 6kV(8kV:option)
Ion-beam diameter 500μm or more (full width at half maximum)
Milling speed 100μm/h(average for 2h, accelerating voltage: 6 kV, Si equivalent value, 100 µm or less from edge)
Maximum specimen size 11 mm (W) 10 mm (L) 2mm (T)
20mm(W)×10mm(L)×5mm(T):optional
Specimen movement range X-axis 10 mm Y-axis 3 mm
Specimen-rotation angle adjustment range ±5°
Specimen-milling swing angle ±30°(patent applied for)
Operation Touch panel
Gas Argon (flow rate controlled by mass-flow controller)
Pressure measurement Penning gauge
Main vacuum pump Turbo-molecular pump
Backing pump Rotary pump
Dimensions and masses Basic unit: 545 mm(W)×550 mm(D)×420 mm(H), 64kg
Rotary pump: 150 mm(W)×427 mm(D)×230 mm(H), 16kg
Optional attachments Accelerating Voltage 8 kV Unit
Specimen Rotation Holder
High Positioning Accuracy CCD Camera
Large Specimen Holder

Installation Requirements

Power Single-phase AC 100 to 120 V 10%, 50/60 Hz, 0.5 to 0.6 kVA
Grounding 100 Ωor less
Argon gas Pressure: 0.15±0.05 MPa (1.0 to 2.0 kg/cm2)
Purity: 99.9999% or more
(argon gas, gas cylinder and regulator should be provided by the customer)
Hose joint JIS B0203, Rc 1/8
Room temperature 15 to 25℃
Humidity 60% or less
  • Specifications subject to change without prior notice.

Application

Application IB-09010CP

More Info

JEOL Instrument Basics

Easy explanation about mechanisms and
applications of JEOL products

Close
Notice

Are you a medical professional or personnel engaged in medical care?

Yes

No

Please be reminded that these pages are not intended to provide the general public with information about the products.

Contacts

JEOL provides a variety of support services to ensure that our customers can use our products with peace of mind.
Please feel free to contact us.