ultra-thin window (UTW) EDS detector
ultra-thin window (UTW) EDS detector
An EDS detector that uses an organic thin film of 0.3 to 0.5 mm thickness as a window material to prevent contaminants onto a detector element.
Aluminum is evaporated on the surface of the organic thin film so as to avoid charging on the window. Compared with the beryllium window EDS detector, the ultra-thin window (UTW) EDS detector better suppresses absorption of low energy X-rays. Thus, the detector can analyze elements lighter than sodium (Na).
Aluminum is evaporated on the surface of the organic thin film so as to avoid charging on the window. Compared with the beryllium window EDS detector, the ultra-thin window (UTW) EDS detector better suppresses absorption of low energy X-rays. Thus, the detector can analyze elements lighter than sodium (Na).
Related Term(s)
Term(s) with "ultra-thin window (UTW) EDS detector" in the description