Analyzes of alkyl phosphonate mixtures

NM230005E

In nuclear magnetic resonance (NMR) analyzes of phosphorus-containing compounds, 1H and 13C signals near phosphorus are affected by heteronuclear spin couplings. Given the natural abundance of 31P at 100%, these splittings are easily visible, reducing sensitivity and increasing complexity. Additionally, some of these compounds are unstable in solvents, leading to the formation of decomposition products, which further complicates the interpretation of 1H and 13C spectra. On the other hand, when analyzing phosphorus-containing samples, 31P NMR information can be invaluable for structural analysis. To illustrate this, we will provide some examples below. In the measurements detailed in this application note, we utilized the ROYAL ProbeTM P+ [1], capable of 1H, 31P, and X triple-resonance measurements. We employed a three-channel 600 MHz NMR spectrometer, the JNM-ECZL600G, featuring one high-frequency (HF) and two low-frequency (LF) channels.

Sample information

Alkyl phosphonates known as Horner-Emmons reagents are used in the olefination reaction. These include diethyl methylphosphonate 1, ethyl dimethyl phosphonoacetate 2, and diethyl (ethoxymethyl) phosphonate 3. They were mixed and dissolved in CDCl3 to create a sample containing 4.8 vol% of each compound.

¹³C NMR spectrum

A conventional 13C spectrum of the sample recorded with 1H decoupling is depicted in Fig. 1. Upon examining the expansions of this spectrum in Fig. 2, an experienced chemist may observe that several signals appear doubled, indicating splitting caused by 13C-31P coupling. However, identifying such signal doublings may prove challenging in the case of unknown samples or very complex mixtures with numerous carbon signals. To illustrate the efficacy of triple-resonance experiments, a 13C spectrum recorded with simultaneous 1H and 31P decoupling is presented in Fig. 2c) and Fig. 2d). A straightforward comparison between the double-resonance and triple-resonance spectra allows for the unequivocal identification of carbon signals split by one-bond M13C-31P couplings, which can exceed 100 Hz in alkyl phosphonates. The signals observed at 10.85 ppm (JCP = 144 Hz)*1, 33.13 ppm (JCP = 135 Hz)*2, and 64.35 ppm JCP = 167 Hz)*3 can thus be attributed to the carbon atoms directly bound to phosphorus, with the one-bond coupling constants provided in parentheses. Smaller splittings are indicative of 2-bond and 3-bond interactions. Table 1 summarizes the 13C chemical shifts and 13C-31P coupling constants.

Fig. 1: ¹³C NMR spectrum
Fig. 2: Expanded regions of  
a) ¹³C{¹H} and b) ¹³C{¹H}{³¹P} spectra in the range of 50-70 ppm,
c) ¹³C{¹H} and d) ¹³C{¹H}{³¹P} spectra in the range of 10-35 ppm
No. 13C/ppm J/Hz
1 10.85 1Jcp=144
2 13.75 -
3 14.57 -
4 16.08 nJcp=7
5 16.15 nJcp=6
6 33.13 1Jcp=135
7 52.84 nJcp=6
No. 13C/ppm J//Hz
8 61.14 nJcp=6
9 61.35 -
10 62.05 nJcp=7
11 64.35 1Jcp=167
12 68.73 nJcp=12
13 165.29 nJcp=6

Table 1: 13C chemical shifts and coupling constants

Difficulty in interpretation of ¹H-³¹P HMBC and use of double-INEPT H-C-P experiment

Fig. 3 shows the1H-31P HMBC spectrum. 1H-31P coupling via 2- and 3-bonds are observed, but it is difficult to distinguish each from this spectrum. On the other hands, by measuring the 1H-13C-31P double-INEPT as shows Fig. 4, only the 2-bonded 1H-31P can be selectively extracted [2]. Then we can distinguish between 2- and 3-bonds by comparing them. the 1JCH and 1JCP couplings are used to obtain correlations in double-INEPT, since 1JCP of alkyl phosphonates have a large coupling constant as mentioned above, it is possible to extract only the H-C-P coupling by setting the parameters according to that.

Correlations observed in 1H-31P HMBC

Fig. 3: ¹H-³¹P HMBC spectrum
Longrange_j : 8 Hz

Correlations observed in 1H-13C-31P double-INEPT

Fig. 4: ¹H-¹³C-³¹P double-INEPT spectrum
j_ch : 140 Hz, j_cp : 145 Hz

Assignment of each ³¹P-¹³C coupling partners

31P-13C correlations can also provide significant insights into the structural analysis of phosphorus-containing compounds. Correctly chosen and optimized 31P-13C correlation experiments enable the identification of 13C coupling partners for each 31P nucleus. Fig. 5 illustrates a 31P-13C LR-HSQC (Long-Range Heteronuclear Single-Quantum Coherence) spectrum. The pulse program employed was a standard HSQC, with parameters optimized for long-range 13C-31P couplings, thereby capturing correlations from weak couplings as well. In Fig.6, we compare the 13C NMR spectrum with three slices extracted from the 31P-13C LR-HSQC spectrum for each phosphorus atom. This facilitates the assignment of 13C signals to individual phosphonates. Additionally, Fig. 6 demonstrates that 31P-13C LR-HSQC enables the detection of carbon atoms up to three bonds away from each phosphorus atom.

Fig. 5: ³¹P- ¹³C LR-HSQC{¹H}{¹³C} spectrum
J_constant : 10 Hz
Fig. 6: Comparison of ¹³C spectrum and slices extracted from ³¹P- ¹³C LRHSQC {¹H}{¹³C} spectrum
※ one-bond correlation

Refference

  • JEOL Application note NM220010E

  • Tetrahedron Letters 48 (2007) 7586–7590

Solutions by field

Related products

JNM-ECZL series FT NMR

Close
Notice

Are you a medical professional or personnel engaged in medical care?

Yes

No

Please be reminded that these pages are not intended to provide the general public with information about the products.

JEOL Instrument Basics

Easy explanation about mechanisms and
applications of JEOL products

Contacts

JEOL provides a variety of support services to ensure that our customers can use our products with peace of mind.
Please feel free to contact us.