Analysis of the aroma components of spearmint essential oil by GC-TOFMS and AI structure analysis

MSTips No. 469

Introduction

Spearmint is a plant in the genus Mentha of the Family Lamiaceae. It has long been known as one of the herbs with a sweet and refreshing aroma. Fresh leaves are often used in cooking, herbal teas, and desserts, and it is also widely used as a flavor for toothpaste and chewing gum. The aroma of spearmint has aromatherapy effect, making it a recommended aroma when you want to energize and refresh a tired body and mind, or when you want to clear your head.

In this MSTips, we introduce the results of analysis of the aroma components of spearmint essential oil by gas chromatography-time-of-flight mass spectrometry (GC-TOFMS). Since essential oils are natural materials extracted from flowers, leaves, and seeds of plants, they are expected to contain substances not registered in the NIST database. Therefore, we used JMS-T2000GC and msFineAnalysis AI, which are capable of AI structural analysis, for the analysis.

Experiment

Commercially available spearmint essential oil was used, diluted by 1/2 with a hexane solution. JMS-T2000 GC AccuTOF™ GC-Alpha (JEOL) was used for GC-MS measurements (Figure 1). The GC-TOFMS measurements were performed using both EI and field ionization (FI) modes with the EI/FI/FD combination ion source. The qualitative data processing was performed with msFineAnalysis AI (JEOL). Measurement conditions are shown in Table 1.

Figure 1

Figure 1 JMS-T2000GC AccuTOF™ GC-Alpha

Table 1 Measurement conditions

GC conditions
Gas Chromatograph 8890 GC (Agilent Technologies)
Column HP-1MS (Agilent Technologies)
30 m x 0.32 mm, 0.25 μm
Inlet 250°C, Split 50:1
Oven temperature 40°C (1 min) - 15°C/min - 250°C (1 min)
Carrier flow He, 1.5 mL/min
MS conditions
Spectrometer JMS-T2000GC (JEOL Ltd.)
Ion source EI/FI combination
Ionization EI (70 eV, 250°C), FI
Mass range m/z 33-800
Analysis software msFineAnalysis AI

Results

Figure 2 shows the TIC chromatogram of spearmint essential oil. Thirty-seven components were detected as having a height threshold of 0.015% or higher. Carvone, which has a sweet and refreshing scent, was detected as the main component, followed by Limonene, which has a sweet-sour and refreshing scent.

Figure 2

Figure 2 TIC chromatograms of spearmint essential oil

Table 2 shows the results of the integrated analysis of the 37 aroma compounds detected in spearmint essential oil. Six of the 37 identified components were not registered in the NIST database, it was able to obtain the compound name and structural formula by AI structural analysis. Figure 3 shows an enlarged TICC around limonene, the second most strongly identified compound. The deconvolution peak detection identified and detected even the low-intensity component as a co-eluting component. The co-eluting component ID010 was also an unknown substance not registered in the NIST library database, but a candidate structural formula could be calculated by the AI structural analysis function as shown below.

Table 2 List of integrated analysis results of spearmint essential oil components

Table 2
Figure 3

Figure 3 TIC chromatograms (left), mass spectrum (center),
AI structural analysis results summary (right) of spearmint essential oil components

Conclusion

GC-TOFMS using JMS-T20000GC and msFineAnalysis AI were used to analyze aroma components in spearmint essential oil. Deconvolution peak detection by msFineAnalysis AI enabled the detection of co-eluted components without missing them. Although some components were not registered in NIST database, it was able to obtain the compound name and structural formula by AI structural analysis. So, it was confirmed that this device and software are effective for analyzing aroma components in naturally derived samples.

 

Solutions by field

Related products

JMS-T2000GC AccuTOF™ GC-Alpha High Performance Gas Chromatograph - Time-of-Flight Mass Spectrometer

msFineAnalysis AI Unknown Compounds Structure Analysis Software

Product category

Close
Notice

Are you a medical professional or personnel engaged in medical care?

Yes

No

Please be reminded that these pages are not intended to provide the general public with information about the products.

JEOL Instrument Basics

Easy explanation about mechanisms and
applications of JEOL products

Contacts

JEOL provides a variety of support services to ensure that our customers can use our products with peace of mind.
Please feel free to contact us.