Comparison of aromatic compounds in two different red wines by using GC-High resolution TOFMS with the integrated data analysis of msFineAnalysis AI

MSTips No.449

Introduction

JMS-T2000GC AccuTOFTM GC-Alpha

It is known that the taste of alcoholic beverages such as  wine has a strong influence from aromatic compounds. It is also known there are many common aromatic compounds that contribute to the typical taste and aromas of wine independent of the production regions, and there are also many aromatic compounds that are strongly dependent on the production regions and the kinds of grapes.

In case of wine, the difference of taste and aroma between the wines that were brewed the different production region is often discussed. It is of interest to know the type of aromatic compounds that contribute to those differences.

The combination of the micro solid-phase extraction (SPME) and gas chromatograph-mass spectrometer (GC-MS) is widely used for the analysis not only of aromatic compounds but also volatile organic compounds. The combined analysis method was used to detect the characteristic compounds of two type of red wines, Bordeaux and Bourgogne. All aromatic samples were measured by not only electron ionization (EI) but also photoionization (PI) to identify the characteristic chemical compounds that were included in the only Bordeaux or only Bourgogne wine. In addition, both measurement data were analyzed by using msFineAnalysis AI software.  The characteristics chemical compounds that differ between the two wines were identified by using the differential analysis functionality of msFineAnalysis AI.

Experiment

Two typical red wines, produced in Bordeaux and Bourgogne, which are generally available in a supermarkets were used as test samples. 2 mL portion of the wine samples were added to 15mL glass vial and  immediately sealed. The SPME fiber was inserted into the headspace of the glass vial and the fiber  was exposed for 20 min at room temperature. All extracted compounds in the SPME fiber were introduced into GC by heating at injection port of GC for 1min. The details of measurement condition were shown in Table 1. To use the differential analysis functionality, each sample was measured for three times using EI ionization. Then those measurement results were applied to msFineAnalysis AI (JEOL Ltd.) together with the measurement results by PI ionization.

 

Table 1. Measurement condition

Results

Total Ion Current Chromatograms (TICCs) obtained as a result of measuring the aromatic components of Bordeaux and Bourgogne wines by EI ionization are shown in Fig.1. In addition, the volcano plot obtained as a result of performing a difference analysis using Bordeaux wine as Sample A (blue) and Bourgogne wine as Sample B (red) is shown in Fig.2. In the volcano plot of msFineAnalysis AI software, the horizontal axis represents the intensity ratio between two samples (Log2(B/A)), and the vertical axis represents statistical reproducibility (-log10(p-value)). By using this volcano plot, it becomes easy to understand the difference in components. The plots in the area with a blue background shown in Fig. 2 are characteristically detected peaks in Sample A, and conversely, those with a red background are the characteristically peaks in Sample B.

 

Fig.1 TICC with EI for Sample A(Blue) and B(Red)

Fig.2 Volcano plot of Bordeaux(Blue) vs. Bourgogne(Red)

 

Table 2.   Integrated qualitative analysis results for the characteristic compounds for Sample A (Bordeaux) and Sample B (Bourgogne)

 

The characteristic compounds in each sample (Bordeaux : Sample A and Bourgogne : Sample B) were identified and listed in Table 2. Totally 14 compounds (9 from Bordeaux, 5 from Bourgogne) were identified. Concerning 8 of 14 compounds, the molecular ions were identified, and the exact m/z values of those molecular ions agreed with the exact m/z values of the estimated compounds. In addition, the EI fragment coverage values for those compounds that are calculated from the relationship between the elemental composition of fragment ions and those of molecular ion were also high values. This means the validity of the compound identification is much higher than by simple library search alone.
It is known that several ‘carboxylic acid esters’ are the typical aromatic compounds that characterized the aroma of wine. Many kinds of methyl and/or ethyl esters of low fatty acids were detected from both of Bordeaux and Bourgogne wine. This means that the esters of low fatty acids strongly contribute the aroma of wines.
The comparison of TICC by EI and PI for Hexanoic acid ethyl ester that was detected from both of Bordeaux and Bourgogne wines is shown in Fig 3. The position of the Hexanoic acid ethyl ester in the volcano plot is also shown in Fig.4. By using the differential analysis functionality of msFineAnalysis, the detailed data analysis and confirming information is available for each peak.

 

Fig.3  TICC of EI and PI for Hexanoic acid ethyl ester detected in Sample A and Sample B

Fig.4  Relative position of Hexanoic acid ethyl ester in Volcano plot

 

Fig.5  TICC of EI and PI for 3,4,4a,5,6,7-hexahydro-1,1,4a-trimethyl-2(1H)-Naphthalenone detected only in Sample B (not in Sample A)

Fig.6  Mass spectrum of 3,4,4a,5,6,7-hexahydro-1,1,4a-trimethyl-2(1H)-Naphthalenone detected in Sample B

 

It is interesting to note that esters of fatty acids with a relatively large number of carbon atoms were detected as the characteristic compounds of Bourgogne wine.
Several compounds such as Propylene Glycol and Methyl salicylate in Bordeaux wines and 3,4,4a,5,6,7-hexahydro-1,1,4a-trimethyl-2(1H)-Naphthalenone in Bourgogne wines were detected as characteristic compounds for only one of the two types of wine.
The TICC’s of EI and PI measurements for 3,4,4a,5,6,7-hexahydro-1,1,4a-trimethyl-2(1H)-Naphthalenone is shown in Fig.5. Although the peak corresponding to 3,4,4a,5,6,7-hexahydro-1,1,4a-trimethyl-2(1H)-Naphthalenone was detected as a shoulder peak beside of a huge peak, it was clearly detected using the chromatographic deconvolution functionality of msFineAnalysis AI. The mass spectra of 3,4,4a,5,6,7-hexahydro-1,1,4a-trimethyl-2(1H)-Naphthalenone measured by EI and PI is shown Fig. 6.

In total 14 compounds were found as characteristic compounds from only Bordeaux and only Bourgogne wine in this study. It is suggested that those compounds contribute the characteristic aroma and taste for both wines.

Conclusion

In this MSTips, we introduced an example of difference analysis by using msFineAnalysis AI concerning aroma components in wines from different production areas.
By using msFineAnalysis AI, qualitative analysis for GC-TOFMS is expected to be more accurate and efficient.

Solutions by field

Related products

JMS-T2000GC AccuTOF™ GC-Alpha High Performance Gas Chromatograph - Time-of-Flight Mass Spectrometer

msFineAnalysis AIUnknown Compounds Structure Analysis Software

Product category

Close
Notice

Are you a medical professional or personnel engaged in medical care?

Yes

No

Please be reminded that these pages are not intended to provide the general public with information about the products.

JEOL Instrument Basics

Easy explanation about mechanisms and
applications of JEOL products

Contacts

JEOL provides a variety of support services to ensure that our customers can use our products with peace of mind.
Please feel free to contact us.