【DISCONTINUED】IB-09020CP Cross Section Specimen Preparation Device

DISCONTINUED

IB-09020CP Cross Section Specimen Preparation Device

This product is no longer available.

If you would like to know the latest information about your preferred product or to find out more about alternatives, please click on the link below. We hope you will continue to use our products.

In addition to the functions of the IB-09010CP, the processing progress can be monitored in a real time using the attached CCD camera.

Features

A monitor camera (x20 to x100), which enables tracking of the processing progress, is mounted.
A function to automatically start the processing without waiting time is included. In this way, the process can be started at the optimal vacuum rate.
There is an optional 8kV power supply available, providing high processing speeds of 300μm/h (on a silicon substrate conversion basis, excrescence quantity : 100μm, 2 hours average)
Easy operation by touch panel

Specifications

Ion accelerating voltage 2 to 6kV
Ion-beam diameter 500μm(full width at half maximum)
Milling speed 100m/h (the average value over 2 hours, at accelerating voltage: 6 kV, specimen: Si, 100 m from edge)
Maximum specimen size 11 mm (width) × 10 mm (length) × 2 mm (thickness)
Specimen movement range X-axis:±10 mm, Y-axis:±3 mm
Specimen-rotation
angle adjustment range
±5°
Specimen-polishing swing angle ±30 (patent applied for)
Operation Touch panel
Gas Argon (flow rate controlled by mass-flow controller)
Pressure measurement Penning gauge
Main vacuum pump Turbo-molecular pump
Dimensions and weights Basic unit: 380mm(W)×570mm(D)×520mm(H), 66kg
Rotary pump: 150mm(W)×427mm(D)×230.5mm(H), 16kg
Optional attachments 8 kV Power Supply
Specimen Rotation Holder(SM-09011)
High Positioning CCD Camera(IB-83910CCD)
Large Specimen Holder

Installation Requirements

Power Single phase 100 to 120 V ±10% AC, 50/60 Hz, 0.5 to 0.6 kVA
Grounding terminal 100Ω or less
Argon gas Pressure: 0.15 ±0.05 MPa (1.0 to 2.0 kg/cm2)
Purity: 99.9999% or more
(argon gas, gas cylinder and regulator should be provided by the customer)
Hose joint: JIS B0203, Rc 1/4
Room temperature 15 to 25℃
Humidity 60% or less
  • Specifications subject to change without prior notice.

Application

Application IB-09020CP

More Info

JEOL Instrument Basics

Easy explanation about mechanisms and
applications of JEOL products

Close
Notice

Are you a medical professional or personnel engaged in medical care?

Yes

No

Please be reminded that these pages are not intended to provide the general public with information about the products.

Contacts

JEOL provides a variety of support services to ensure that our customers can use our products with peace of mind.
Please feel free to contact us.