Analysis of di(2-ethylhexyl)phthalate, dichloroacetonitrile and chloral hydrate by GC-MS method using nitrogen carrier gas

MSTips No.398

1. Overview

JMS-Q1600GC UltraQuad™ SQ-Zeta

Helium (He), which is widely used as a carrier gas for GC, can have problems such as temporary price increases and unstable supply conditions due to various reasons, and when supply delays occur, it is necessary to consider using another type of carrier gas as an alternative. Hydrogen and nitrogen are the main alternative gases being considered, with nitrogen gas being relatively easier to introduce into GC-MS when safety is a priority. In this report, di(2-ethylhexyl)phthalate, dichloroacetonitrile, and chloral hydrate, which are water quality control target items in water quality testing, were measured by the same column using nitrogen carrier gas. (For He, see MSTips No.325.) As a result, good linearity of the calibration curve and reproducibility at the lower limit of quantification were obtained for all compounds, and are presented in this report.

1.1. Measurement Condition

The measurements were performed using a gas chromatograph quadrupole mass spectrometer "JMS-Q1600GC UltraQuad™ SQ-Zeta". The measurement conditions are shown in Table 1. As mentioned above, the column used was the same, and each compound was measured by changing the GC conditions. The monitor ions for SIM described in the Japanese "Notification Law" were used.

Table. 1 Measurement Condition of Each Compound

Parameter Value
Di(2-ethylhexyl)phthalate Dichloroacetonitrile & Chloral hydrate
GC Oven temp. 50°C (2 min)→20°C/min→180°C (0 min)→
5°C/min→260°C (10.5 min)→10°C/min→
280°C (5 min), Total 42 min
35°C (3.5 min)→15°C/min→100°C (0 min)→
20°C/min→250°C (3 min),
Total 18.3 min
Column flow (Nitrogen) 1 mL/min 1 mL/min
Injection mode Splitless, Purge time 1 min Splitless, Purge time 0.5 min
Injection volume 2 uL 1 uL
Column GL Sciences Inc. InertCap 1 MS, 3 m x 0.25 mm id, 1 μm film thickness
Inlet temp. 250°C
MS Interface temp. 250°C
Ion source temp. 250°C
Ionization EI (20 eV, 50 μA)
Acquisition mode SIM

2. Di(2-ethylhexyl)phthalate

2.1. Measurement

The test solution was diluted with n-hexane to achieve concentrations of 5, 10, 15, and 20 μg/L of di(2-ethylhexyl)phthalate in the test water before pretreatment. Phenanthrene-D10 was added as an internal standard at a concentration of 250 μg/L.

2.2 Result

The calibration curve for di(2-ethylhexyl)phthalate is shown in Figure 1. The coefficient of correlation for the relative calibration curve was more than 0.999. SIM chromatogram and the coefficient of variation of the quantitation value for the 5 μg/L sample measured continuously at n=5 are shown in Figure 2 and Table 2, respectively. The coefficient of variation of the quantification value at 5 μg/L, which is less than 1/10 of the target value of 80 μg/L, was within 5%, a good result.

Figure 1

Figure 1. Calibration curve of Di(2-ethylhexyl)phthalate

Figure 2

Figure 2. SIM chromatogram of Di(2-ethylhexyl)phthalate at 5 μg/L

Table 2. Coefficient variation (C.V.) of Di(2-ethylhexyl)phthalate at 5 μg/L

Quantitation value (μg/L) C.V. %
#1 #2 #3 #4 #5
5.16 5.19 5.19 5.18 5.18 0.2

3. Dichloroacetonitrile and Chloral hydrate

3.1. Measurement

The test solution was prepared by dilution with MTBE to achieve concentrations of 1, 3, 5, and 15 μg/L of dichloroacetonitrile and chloral hydrate in the test water before pretreatment. 1, 2, 3-trichloropropane was added as an internal standard at a concentration of 125 μg/L.

3.2. Result

The calibration curves for dichloroacetonitrile and chloral hydrate are shown in Figure 3. The correlation coefficients for the relative calibration curves were more than 0.999. The SIM chromatograms and the coefficient of variation values of the quantification values for the continuous measurement of 1 μg/L samples at n=5 are shown in Figure 4 and Table 3, respectively. The coefficient of variation of the quantification value at 1 μg/L, which is less than 1/10 of the target value (dichloroacetonitrile: 10 μg/L, chloral hydrate: 20 μg/L), was within 5%, a good result.

Figure 3 Dichloroacetonitrile
Figure 3 Chloral hydrate

Figure 3. Calibration curve of dichloroacetonitrile & chloral hydrate

Figure 4 Dichloroacetonitrile
Figure 4 Chloral hydrate

Figure 4. SIM chromatograms of dichloroacetonitrile & chloral hydrate at 1 μg/L

Table 3. Coefficient variation of dichloroacetonitrile & chloral hydrate at 1 μg/L

Compound name Quantitation value (μg/L) C.V. %
#1 #2 #3 #4 #5
Dichloroacetonitrile 0.93 0.92 0.95 0.93 0.91 1.9
Chloralhydrate 0.99 1.03 1.01 1.00 1.00 1.7

4. Summary

Di(2-ethylhexyl) phthalate, dichloroacetonitrile, and chloral hydrate were measured by the same column using a nitrogen carrier gas. The linearity of the calibration curve and the reproducibility at concentrations less than 1/10 of the target value were good, confirming that the measurement by the same column using a nitrogen carrier gas can be used adequately.

 

Solutions by field

Related products

JMS-Q1600GC UltraQuad™ SQ-Zeta Gas Chromatograph Quadrupole Mass Spectrometer

Close
Notice

Are you a medical professional or personnel engaged in medical care?

Yes

No

Please be reminded that these pages are not intended to provide the general public with information about the products.

JEOL Instrument Basics

Easy explanation about mechanisms and
applications of JEOL products

Contacts

JEOL provides a variety of support services to ensure that our customers can use our products with peace of mind.
Please feel free to contact us.