Delta Tips: Baseline Correction in 2D Spectra

NMDT_0029

If baseline is distorted in a 2D spectrum, bands of noise may appear along F1- and/or F2-axis. Such noise complicates the analysis of low-intensity peaks and is a serious problem in 2D NOESY and ROESY spectra. The figures below demonstrate automatic baseline correction along F2-axis.

Select Actions – Base Correct – Transpose 0 to apply automatic baseline correction in F2-domain. There are no parameters to set.

Delta Tips
∗ To apply baseline correction in F1-domain, select Transpose 1.