Is the sample dissolved uniformly? Insufficient dissolution reduces resolution

NM200006E

As shown below, non-uniform sample concentrations cause local magnetic susceptibility distortion. Therefore, magnetic field adjustment is difficult, and good resolution spectra cannot be sometimes obtained. Please keep in mind that the sample concentration may be uneven, even if the sample appears to be dissolved.

uniform concentratoin

Example of Menthol

Here, a simple example is shown with menthol solution. As we know, menthol is extremely soluble compound with methanol, and we would imagine it is very easy to dissolve completely. Menthol powder was placed in a NMR sample tube, and deuterated methanol was added to make a 10% solution. After NMR sample tube was shaking back and forth several times, sample solution became transparent as shown in the photo below. Figure A shows a spectrum obtained by measuring this sample, which has poor resolution. Subsequently, this sample tube was ejected from the magnet, and shaken well with same way. Figure B shows the spectrum of this sample after second well shaking, with a slight improvement in resolution, but still not enough. When this sample tube was shaken vigorously up and down finally, the resolution improved dramatically, as shown in Figure C. Even for easily soluble samples, it is important for good resolution spectra to completely dissolve NMR sample with solvent.

10% Menthol
shake it firmly until a uniform concentration is achieved
Please see the PDF file for the additional information.
Another window opens when you click.

PDF 184.8KB

SEARCH APPLICATIONS

Close
Notice

Are you a medical professional or personnel engaged in medical care?

Yes

No

Please be reminded that these pages are not intended to provide the general public with information about the products.

JEOL Instrument Basics

Easy explanation about mechanisms and
applications of JEOL products

Contacts

JEOL provides a variety of support services to ensure that our customers can use our products with peace of mind.
Please feel free to contact us.