Novel Three-Dimensional Correlation Experiments to Reveal Local Proton Networks under Fast Magic-Angle Spinning at Natural 13C Abundance 

NM170009

1H double quantum (DQ)/1H single quantum (SQ) correlation solid-state NMR spectroscopy is widely used to obtain internuclear 1H-1H proximities, especially at fast magic angle spinning (MAS) rate > 60 kHz. However, 1H signals are not well-resolved due to intense 1H-1H homonuclear dipolar interactions even at the attainable maximum MAS frequencies of ~100 kHz to date. We have recently introduced novel three-dimensional (3D) experiments to resolve the 1H DQ/1H SQ correlation peaks using the additional 13C dimension. Although the low natural abundance of 13C (1.1%) significantly reduces the sensitivities, the 1H indirect measurements alleviate this issue and make this experiment possible. The two different implementations of 13C/1H DQ/1H SQ correlations and 1H DQ/13C/1H SQ correlations have been discussed and demonstrated using L-histidine.HCl.H2O at natural abundance to reveal the local 1H-1H networks nearby each 13C. In addition, the complete 1H resonance assignments have been achieved from a single 3D 13C/1H DQ/1H SQ experiment. We have also demonstrated the applicability of our proposed method on a biologically relevant molecule capsaicin.

The 2D 1H DQ/1H SQ spectrum of capsaicin (left) is too complex due to signal overlaps. On the other hand, the 13C-filtered spectrum (right) yields well-resolved peaks and thus signal assignment is possible. Both spectra were recorded at 70 kHz MAS. Please refer to the original paper for more detail.

Reference

  • M. Malon, M. K. Pandey and Y. Nishiyama, J. Phys. Chem. B, 2017, 121, 8123-8131.
If you want to see a printed version,
please click this PDF file.

PDF 266KB
Close
Notice

Are you a medical professional or personnel engaged in medical care?

Yes

No

Please be reminded that these pages are not intended to provide the general public with information about the products.

JEOL Instrument Basics

Easy explanation about mechanisms and
applications of JEOL products

Contacts

JEOL provides a variety of support services to ensure that our customers can use our products with peace of mind.
Please feel free to contact us.