Single-Tuned 4mm Cryocoil MAS Probe
NM140018E
A tunable single-tuned 4 mm Cryocoil MAS probe was developed, and demonstrated for 6Li and 29Si under 14.1 T WB magnet. The sensitivity enhancement factor attained by the Cryocoil MAS probe is 4.4 times compared with the same experimental condition using the conventional probe.
Typical Specification:
- Detection Coil Temperature: 14 K
- Sample Temperature: 273 - 353 K
- Rotor O.D.: 4 mm (volume: 50 ul)
- Spinning Speed: 18 kHz max.
- Tuning Range: 73 - 192 MHz
- Input Power: 250 W max.
Sensitivity and RF field strength enhancement by Cryocoil MAS Probe
Reference
T. Mizuno, K. Hioka, K. Fujioka, K. Takegoshi, Rev. Sci. Instrum. 79 (2008), 044706.
T. Mizuno, K. Takegoshi, Rev. Sci. Instrum. 80 (2009), 124702.