A new field of solid-state NMR by very fast MAS:1H DQMAS NMR for volume limited samples

NM100005E

1H DQMAS NMR at a very fast MAS frequency of 80 kHz is presented. 1H DQMAS spectra give information on proximity between protons. 1H-1H homonuclear dipolar decoupling has been applied to 1H DQMAS to enhance resolution, i.e. 1H DQ-CRAMPS (CPL 394 (2004) 423, JACS 126 (2004) 13230, JEOL application note: NM090008E). Here we show that highly resolved 1H DQMAS spectra can be obtained at a very fast MAS rate of 80 kHz without 1H-1H decoupling. This can be done by 1 mm MAS probe which we recently developed. 1H DQMAS by 1 mm MAS probe realizes high sensitivity collection of volume limited samples of 0.8 uL due to 1H observation without rf irradiation and high sensitivity per unit volume.

DQ excitation and reconversion are done by post C7 under moderate MAS rate. Here we applied high efficiency and recoupling scheme R1252, which is developed by Wang et al. This realizes efficient observation with low rf field strength requirements. The sequence is shown below.

This work has been supported by JST-SENTAN(JAPAN).

Referance

Q. Wang, B. Hu, O. Lafon, J. Trebosc, F. Deng, J.-P. Amoureux, J. Magn. Reson. 203 (2010) 113-128.

SEARCH APPLICATIONS

Close
Notice

Are you a medical professional or personnel engaged in medical care?

Yes

No

Please be reminded that these pages are not intended to provide the general public with information about the products.

JEOL Instrument Basics

Easy explanation about mechanisms and
applications of JEOL products

Contacts

JEOL provides a variety of support services to ensure that our customers can use our products with peace of mind.
Please feel free to contact us.