A new field of solid-state NMR by very fast MAS:sub-micro-litter sample 14N NMR collection in a few minutes

NM100003E

Rapid observations of 14N NMR for sub-micro-litter samples are presented. 14N is  rarely observed by NMR despite its high natural abundance (99.6%) and its  importance in bio and material sciences. The difficulties in observing 14N resonances  are due to its integer spin quantum number (I=1) and quadrupolar interactions.  Because of these difficulties 15N NMR with expensive 15N isotopic-labeling is usually  applied. Very fast MAS by 1 mm MAS system, which we recently developed, realizes  quick observation of 14N NMR resonances of sub-micro liter samples within a few  minutes, combining 1H-14N solid-state HMQC methods. The 1H-14N D-HMQC of dipeptide (glycyl-L-alanine) of 0.8 uL gives clearly separated peaks of two distinct  nitrogens. Slightly different 1H shifts for each peak can also be obtained. No isotopic  labeling methods are applied to the sample.

This work has been supported by JST-SENTAN(JAPAN).