Make it possible to measure a small amount of fluorine samples - 19F MAS probe -

NM090007E

19F MAS probe is recommended for the users who want to:

 ・Observe small 19F signals;
 ・Reduce spinning sidebands;
 ・Improve quantitative evaluation,

  in 19F solid-state NMR spectroscopy.
 



 

Three advantages of 19F MAS probe

#1. Significant reduction of 19F background signals.
Weak signals can be observed.
#2. Fast spinning. (2.5mm: 35kHz, 3.2mm: 24kHz, 4mm: 19kHz)
Spinnning sidebands are greatly reduced.
#3. Strong RF irradiation.
Wide frequency ranges are uniformly excited.

CPMAS vs. 19F MAS probes

19F MAS spectra of a grain of fluorine rubber observed using CPMAS probe (left) and 19F MAS probe (right).

19F background signals
 become 1/100.



 

Effects of fast spinning

19F MAS spectra of PCTFE observed at the spinning frequencies of 24kHz (upper) and 10kHz (lower).

Spinning sidebands
 are greatly reduced.