H2BC & Edited H2BC Selective observation of 2JCH

NM070003E

Long-range correlation peaks in HMBC measurements appear depending on the magnetization transfer time 1/2 nJCH. However, the value of nJCH is not deter-mined by the bond number, and so the correlation peaks give no information on  the bond number. In particular, the close values of 2JCH and 3JCH make it difficult to discriminate the corresponding correlations. H2BC (Heteronuclear 2-Bond Correlation) detects only two-bond away correlations, clarifying the assignments. Furthermore, multiplicity edited H2BC may discriminate attached carbon numbers.
strychnineのEdited H2BC スペクトル
Edited H2BC spectrum of strychnine
1J max = 170 Hz, 1J   min = 125 Hz, y_points = 128, T = 22 msec

H2BC and HSQC spectra of strychnine.
H2BC and HSQC spectra of strychnine.
strychnine
strychnine

By linking correlation peaks in H2BC and HSQC spectra, 13C connectivities can be revealed. Thus, the combination of H2BC and HSQC gives a method for determining sequences of 13C bonding with 1H.

Pulse diagram of H2BC and Edited H2BC measurements.
Pulse diagram of H2BC and Edited H2BC measurements.

Reference

N.T. Nyberg, J.Ø. Duus and O.W. Sørensen, J. Am. Chem. Soc., 127, 6154 (2005)
N.T. Nyberg, J.Ø. Duus and O.W. Sørensen, Magn. Reson. Chem., 43, 971 (2005)

SEARCH APPLICATIONS

Solutions by field

Close
Notice

Are you a medical professional or personnel engaged in medical care?

Yes

No

Please be reminded that these pages are not intended to provide the general public with information about the products.

JEOL Instrument Basics

Easy explanation about mechanisms and
applications of JEOL products

Contacts

JEOL provides a variety of support services to ensure that our customers can use our products with peace of mind.
Please feel free to contact us.