IMPEACH-MBC ~Improvement of wide nJCH range 1H-detected long-range shift correlation measurements~

NM060013E

In HMBC measurements, signal intensities of long-range correlation peaks depend on magnetization transfer time 1/2 nJCH. A basic HMBC pulse sequence, assuming a fixed value of nJCH, cannot give correlation peaks derived from various nJCH values at once (Fig.3 top). ACCORD-HMBC, incorporating variable delays, acquires a wider range of nJCH orrelations, but its spectrum may distort along the indirect axis (Fig.3 middle). IMPEACH-MBC (IMproved PErformance ACcordion HMBC) solves this problem, by adding cancelling inverse variable delays to ACCORD-HMBC to generate a constant-time sequence. IMPEACH-MBC measurements give spectra where long-range shift correlations can easily be analysed (Fig.3 bottom).

Pulse diagram of IMPEACH-MBC measurements.

Fig.1 Pulse diagram of IMPEACH-MBC measurements.

τ1 = 1/2{1JCHmax + 0.146(1JCHmax – 1JCHmin)}–1, τ2 = 1/2{1JCHmax – 0.146(1JCHmax – 1JCHmin)}–1
vd = 1/2(nJCHmin)–1 – y_point[{1/2(nJCHmin)–1 – 1/2(nJCHmax)–1}/(y_points – 1)]
D = initial_delay* + y_point[{1/2(nJCHmin)–1 – 1/2(nJCHmax)–1}/(y_points – 1)]
*initial_delay = 2(PFG + recovery time)

4IMPEACH-MBC spectrum of 4-methyl umbelliferone.

Fig.2 IMPEACH-MBC spectrum of 4-methyl umbelliferone.
nJCHmax = 10 Hz, nJCHmin = 2 Hz, 1JCHmax = 170 Hz, 1JCHmin = 120 Hz, y_points = 512

Spectrometer:JNM-ECX400

HMBC

hmbc_pfg.ex2
long_range_j = 8 Hz
j_constant = 140 Hz

HMBC

↓ cover a wide range of nJCH

ACCORD-HMBC

hmbc_accord_dec_pfg.ex2
y_points = 256
lr_j_min = 1 Hz
lr_j_max = 16 Hz
j_ch_min = 120 Hz
j_ch_max = 160 Hz

ACCORD-HMBC

↓ remove JHH-modulation

IMPEACH-MBC

impeach_mbc_dec_pfg.ex2
y_points = 512
lr_j_min = 2 Hz
lr_j_max = 10 Hz
j_ch_min = 120 Hz
j_ch_max = 170 Hz

IMPEACH-MBC
Spectrometer:JNM-ECX400

Reference

C.E. Hadden, G.E. Martin and V.V. Krishnamurthy, J. Magn. Reson., 140, 274 (1999)

SEARCH APPLICATIONS

Close
Notice

Are you a medical professional or personnel engaged in medical care?

Yes

No

Please be reminded that these pages are not intended to provide the general public with information about the products.

JEOL Instrument Basics

Easy explanation about mechanisms and
applications of JEOL products

Contacts

JEOL provides a variety of support services to ensure that our customers can use our products with peace of mind.
Please feel free to contact us.