Qudruple-resonance probe: C5HPD/FG

NM060012E

Example 1:13C spectra of ATP



Sample:ATP/CDCl3
Spectrometer:
JNM-ECA600
Probe:C5HPD/FG

Example 2:13C-31P shift-correlation spectrum of BINAP

13C-31P shift-correlation spectrum(w/ 1H decoupling)

Sample:
(R)-(+)-BINAP/CDCl3
Spectrometer:
JNM-ECA600
Probe:C5HPD/FG

A C5HPD/FG probe is a quadruple-resonance probe which can irradiate RF pulses for 13C, 1H, 31P, and 2H at once. For phosphorus-containing compounds, 13C signals may split due to couplings with adjacent 31P nuclei, and decoupling for 31P as well as 1H using a C5HPD/FG probe makes 13C spectra simple (Example 1). Also, this probe allows for measurements of 13C-31P correlation spectra (Example 2). Thus, a C5HPD/FG probe is useful for analyzing phosphorus-containing compounds.