Low dew point membrane type air drier

NM050020E


Fig. 1  Conventional standard air drier built-in compressor in use.


Fig. 2  Low dew point membrane type air drier (dew point: -40℃) in use.
※ Conventional compressor is concurrently used.
Sample: deuterated benzene 
Spectrometer: JNM-ECA500 
Probe: TH5AT/FG2 
Accumulation: 512 scan

Among NMR instruments, an air drier is used to generate dried gas for tempera-  ture control, sample spinning, sample loading/ejecting, and so on. In the air, some  moisture may remain depending on the drying performance, yielding unwanted  vapor signals especially in the variable temperature experiments. The air drier of  the standard configuration has sufficient capability for usual measurements to  suppress the unwanted signal at the level lower than 13 C satellite signals and  impurity signals of a small amount of normal solvent (less than 2%) contained in  commercial deuterated solvent (Fig. 1). However, the unwanted signals may be an  obstacle when the small signals in the vapor region are quantitatively observed. In  such cases, vapor signals can be reduced to the undetectable level, by using a low  dew point membrane air drier having higher drying capability (Fig. 2).

※ To use the low dew point membrane air drier, a high pressure compressor may separately be  required to increase gas flow.

SEARCH APPLICATIONS

Close
Notice

Are you a medical professional or personnel engaged in medical care?

Yes

No

Please be reminded that these pages are not intended to provide the general public with information about the products.

JEOL Instrument Basics

Easy explanation about mechanisms and
applications of JEOL products

Contacts

JEOL provides a variety of support services to ensure that our customers can use our products with peace of mind.
Please feel free to contact us.