Measurement of Polymer Intermediates by Time Resolved ESR II

ER100001E

Time Resolved ESR (ES-CIDEP3 - TR-ESR) can measure short-lived radicals that arise from the very early stages of photochemical reactions with ~ 200 ns time resolution. (Refer to Application note ER090004).

Fig.1 Radical Polymerization Scheme Initiated by Photochemical Reaction
Fig.1 Radical Polymerization Scheme Initiated by Photochemical Reaction

Fig.1 shows the radical polymerization reaction of vinyl monomer using TMDPO as polymerization initiator. As DPO・ arising from laser decomposition of TMDPO has a very high reactivity, addition reaction proceeds with the monomer.
The starting radical, arising during the initial polymerization process using ethylvinylether as monomer, was measured by ES-CIDEP3 and the result is shown in Fig. 2.
The spectrum when the intensity of the starting radical grows and decays in a very short time scale has been observed. The time resolved ESR spectrum just 2 ms after laser irradiation is shown in Fig. 3. The hfs parameters, ( aP = 6.77 mT, aHα =1.48 mT, aHβ= 1.56 mT, aHγ = 0.18 mT) correspond well with the structure of the starting radical.
As noted above, TR-ESR can be a very useful tool for analysing the reaction in photochemical processes.

Fig.2 2D Spectrum of Starting Radical of Ethylvinylether
Fig.2 2D Spectrum of Starting Radical of Ethylvinylether
Fig3. Spectrum of Starting Radical of Ethylvinylether
Fig3. Spectrum of Starting Radical of Ethylvinylether

Reference

Y.Mizuta et.al. Appl. Magn.Reson. 19, 93-110 (2000)

SEARCH APPLICATIONS

Close
Notice

Are you a medical professional or personnel engaged in medical care?

Yes

No

Please be reminded that these pages are not intended to provide the general public with information about the products.

JEOL Instrument Basics

Easy explanation about mechanisms and
applications of JEOL products

Contacts

JEOL provides a variety of support services to ensure that our customers can use our products with peace of mind.
Please feel free to contact us.