Direct analysis of stains on cloth caused by cosmetic

MSTips No.D023 Application Note for DART

 Foundation is one of the essential cosmetics for women. At the mean time, it often makes a stain on clothing. We directly analyzed foundation stains on cloth (Fig. 1) without sample preparation by using DART (Direct Analysis in Real Time) ion source installed on AccuTOF™ LC  series. Foundation from 3 manufacturers were measured and compared.

Helium was used as the DART gas. The DART gas heater was set to 200 °C. A small piece of cloth on which foundation was deposited was presented to the DART ion source.
Ions at m/z 371 were observed from the foundation of all 3 manufacturers. Based on the measured accurate masses, the elemental formula of the ions was elucidated to be C10H31O5Si5 and inferred to be the [M+H]+ ions of decamethylcyclopentasiloxane. Ions at m/z 223 were found only from Bland A and their elemental composition was elucidated to be C10H15O2Si2 from the measured accurate m/z.
By using DART ion source, the nature and cause of stains on cloth can be determined without sample preparation.

Fig. 1  A foundation stain on a piece of cloth.
Fig. 1  A foundation stain on a piece of cloth.

Fig. 2  Mass spectra of foundation stains (Top: Bland A, Middle: Bland B, Bottom: Bland C)
Fig. 2  Mass spectra of foundation stains (Top: Bland A, Middle: Bland B, Bottom: Bland C)

Measured m/z Calculated m/z Error (x 10-3) Elemental composition Unsaturation
223.06194 223.06106  0.88 C10H15O2Si2 5.5
371.10100 371.10178 -0.78 C10H31O5Si5 0.5

Table 1 Elucidated elemental compositions of the ions found in the mass spectrum of Bland A.

Please see the PDF file for the additional information.
Another window opens when you click.

PDF 541KB

Related Products

Close
Notice

Are you a medical professional or personnel engaged in medical care?

Yes

No

Please be reminded that these pages are not intended to provide the general public with information about the products.

JEOL Instrument Basics

Easy explanation about mechanisms and
applications of JEOL products

Contacts

JEOL provides a variety of support services to ensure that our customers can use our products with peace of mind.
Please feel free to contact us.