Analysis of polydimethylsiloxanes by GC/TOFMS

MS Tips No.131 The author would like to thank Mr.Junji Takaie of Showa Denko K.K. for his generous gift of the samples.

Three polydimethylsiloxane compounds; octamethylcyclotetrasiloxane (sample No. 1), decamethylcyclopentasiloxane (No. 2) and octa(dimethylsiloxy)silsesquioxane (No. 3), were analyzed using JMS-T100GC. The samples No. 1 and No. 2 were relatively small molecules with molecular weights of 296 and 370 respectively and could be introduced through the mass reference sample inlet (reservoir.) They can be used as internal mass references for accurate mass measurement. The sample No. 3 has a molecular weight of 1,016 and was difficult to be introduced through the reservoir. Sample No. 3 was analyzed by GC/MS.
Please see the PDF file for the additional information.
Another window opens when you click.

PDF 162KB