Features
Carryover Avoidance
Carryover avoidance mechanism is ready for various sample types. Flexible-Access technology and 3 wash modes boosts throughput and assures high-quality analysis.
Probe Wash

Mixer Unit Wash

Cuvette Wash

New Dilution Cuvettes & Drying Nozzle

Micro-volume Technology
The unique dilution mechanism, the strong tradition of BioMajesty™ series, bolstered by the state-of-the-art technologies; The chemistry analyzer contributes to advanced quality healthcare by delivering "more data from the single droplet" to the scene of medical practice.

Small Volume Sample Mode

Sample Cup Adapter for Small Volume Sample

Proven Micro-Volume Technology - Extra micro volume measurement with Min. 40μL reaction volume
Innovative Operation
Innovative 4 Samplers with advanced user interface bring distinguished operability, further accelerating the throughput of your laboratory.
REAGENT PAUSE Function
Reagent can be refilled or replaced without cancelling the operation. Selectable of running re-calibration and QC sample measurement after resuming from the pause.Reagent Bottle Switching
A calibration curve can be created per reagent pair with quick access to each control data.ISE PAUSE Function
ISE reagent can be replaced while running time-critical ISE tests. Once ISE reagents are replaced, measurement quickly resumes after priming.Automatic Reagent Blank Measurement
Reagent blank measurement can be configured to run automatically at system startup. It requires no manual intervention such as setting saline for the blank measurement.Active Trace Feature
Information vital to data analysis is promptly accessible from on-screen test results. Reaction Process Data Check Assistance Tool captures abnormalities in real-time.
Simple Maintenance
Regular maintenance area is allocated within a 40 cm reach from the front. Easy maintenance requires no tool or no manual repositioning in replacing the cuvettes, the probes, the mixing rods, the lamp and sealing material for pumps.

Specifications
Model name | Automatic Analyzer JCA-ZS050 |
---|---|
Measurement method | Open discrete Reagent Pickup type
Pre dilution system |
Throughput | Throughput: Maximum 1,800 tests/hour
Photometric: Maximum 1,200 tests/hour ISE: Maximum 600 tests/hour |
Simultaneous measurements | Maximum 60 assays (63 assays with ISE) |
Analysis methods | Photometric method, reaction rate assay (RRA), CRA method, IMA method and more |
Sampling | |
- Sample Tray | STT: 80 sample positions with Built-in Barcode Reader
CTT: 60 Refrigerated sample positions |
- LAS | Point-in-space Sampling for LAS |
- Sample dispensing volume | 2.0 to 30 μL(0.1 μL / step) |
- Sample volume :(after dilution) | 1.0 to 25 μL(0.1 μL / step) |
- Sensors | Liquid-level sensor, Clot sensor, Clash sensor |
Reagent | |
- Reagent Tray | 2 trays refrigerated with Built-in Barcode Readers |
- Reagent container | 20 mL, 40 mL and 70 mL |
- Reagent volume/assay | 5.0 to 70 μL (0.1 μL /step) |
Reaction Carousel | |
- Reaction cuvette | Plastic, light-path length: 4 mm |
- Reaction volume | 40 to 120 μL |
- Reaction time | 3, 4, 5, or 10 minutes (if extended, 15 or 21 minutes) |
- Reaction temperature | 37 ± 0.1 °C |
- Reaction bath | Temperature-controlled reaction bath with inert oil |
- Mixer | SSR-Spin Mixer |
Measurement method | |
- Measurement points | Maximum 65 points |
- Measurement wavelengths | 340 to 884 nm (14 wavelengths), 1 or 2 wavelengths for calculation |
Maintenance | Automatic startup and shutdown timer |
Workstation | PC, LCD color monitor |
Host interface | RS232C |
Size | Dimensions: 1,470 (W) × 904 (D) × 1,493 (H) mm, Weight: 550 kg |
Rack Handler (Optional)
Model name | RH100 |
---|---|
Sample loading | 100 samples (20 racks) |
Sample unloading | 100 samples (20 racks) |
STAT sample | Loadable from the STAT lane
Self-starting by simply placing the rack on the lane |
Rack types | Per sample category
(Patient sample, calibrator, and QC sample) |
Barcode | |
Size | Dimension: 800 (W) × 1,025 (D) × 1,275 (H) mm, Weight: 230 kg |

Application
Application JCA-ZS050
Gallery

-
JCA-ZS050
-
JCA-ZS050+RA101
-
JCA-ZS050+RH100
More Info