Features
This is used to control the sample temperature during measurement in the range from -170°C to 200°C. By improving the response of the heater, very smooth temperature control with improved stability is achieved. Compatible with photo-irradiation attachment.
Specifications
Variable temperature range | -170 to 200°C |
---|---|
Temperature stability | ± 0.5 °C |
Temperature control | Automatic PID method |
Metal dewar capacity 1) | 10 L |
Compressor flow rate | 17 L / min (Air) |
Power supply | AC 100 V, 5.5 A (incl. Compressor) |
Applicable cavity | ES-UCX2 / MCX1A / MCX3B / ODMRC / 14040DMC |
Sample tube | 5 mm OD |
1) A 30 L dewar is available as an option. (Lowest temperature: -160°C)
Composition
Temperature controller
Heat blower tube
Metal dewar
Flow sensor
Air compressor
VT adapter
Dewar adapter
Cooling water dust remover
Application
Application ES-13060DVT5
Variation of ESR line shape with temperature
Long-term measurement using variable temperature controller (ES-13060DVT5)
More Info